Solicitation
SP-300 Potentiostat, Potentiostat/galvanostat board and a standard cell cable with electrometer
RFQ371738
Department of Energy, ORNL Ut-Battelle LLC-DOE Contractor. Analytical Laboratory Instrument Manufacturing.
Response deadline
May 7, 2025 at 5:00 PM EDT
Closed 567 days ago. Posted April 29, 2025. Scheduled to archive May 10, 2025.
Get the next one like this from Department of Energy by email, the day it is posted.Find my opportunities
This one closed in 2025
59 like it are open now in analytical laboratory instrument manufacturing
- Adjustable-Volume Laboratory Micropipettes and Pipettors
Department of Energy, ORNL Ut-Battelle LLC-DOE Contractor
Combined synopsis and solicitationNAICS 334516Ship to Oak Ridge, TNPR519028Due todayNov 25 - ARBITRARY WAVEFORM GENERATOR
Department of the Navy, NSWC Crane
Combined synopsis and solicitationNAICS 334516Ship to Crane, INN0016427Q0014Due todayNov 25 - Brand Name or Equal Teledyne Instruments Liquid Sy
Department of Energy, National Energy Technology Laboratory
SolicitationSmall businessNAICS 334516Ship to the buyer89243327QEX000002Due todayNov 25
All 59 open in analytical laboratory instrument manufacturing
Reading this notice
Worked out from the publications and the award on this page.
- Kind of notice
- Solicitation. A formal request for proposals, quotes or bids. Offers are due by the response deadline. Glossary
- Who may compete
- SAM.gov lists no set-aside on this notice, so it is not reserved for small businesses. Small, under NAICS 334516 (analytical laboratory instrument manufacturing), means no more than 1,000 employees, counted with affiliates. What a set-aside is
- Time to respond
- Bidders had 8 days from the day the solicitation was posted, April 29, 2025, to the deadline.
Size standard from 13 CFR 121.201. The solicitation itself is what binds: check the code and standard it states.
Description
As published on SAM.gov.
SP-300 Potentiostat, Potentiostat/galvanostat board and a standard cell cable with electrometer
Request for the exact or domestic equivalent multi-channel, modular design potentiostat/galvanostat/EIS analyzer based on a portable chassis is required. Needs to be capable of USB and ethernet communication. Needs the following operational modes: floating, grounded, counter electrode to ground, working electrode to ground. EIS experiment functionality from 10 micro-Hz to 7 MHz. No separate analyzer should be required. Both sine and multi-sine techniques need to be available. Current and voltage should be possible to be measured simultaneously, with compliance voltage of +/- 12 Volt and reference voltage adjustable between -10 to +10 volts, with same range between working and reference electrodes. Resolution must be programmable from 300 micro-volts down to 1 micro-volts. Needs to apply and measure currents from 1 pico-amp to 500 milli-amps, a current control resolution of 0.004% and accuracies to 0.1% of FSR. Should be possible to be interfaced with boosters of 1,2,4 or 10 Amps on the same chassis. Multi-amp boosters should be possible to be paired for higher currents. Should be expandable to include a linear scan generator for true analog waveforms, up to 1 MV/s with data aquisition rates as fast as 1 micro-second. Requires an ultra-low current option to +/- 100 femto-amps on 1 pico-amp range with greater than 10^14 ohms input impedance. Should be capable of switching control modes in under 5 microseconds, without adverse spiking of the current in the connected cell. Requires an external input for control of potential or current via external analog input from separate device. Needs to have at least nine user-selectable bandwidths for full stability control on channels. Analog filters (50 kHz, 1 kHz, 5 Hz) as well as digital filtering in software. Needs to be able to acquire data at 20 micro-second frequency. Needs to be capable of 2-5 terminal measurements (for simultaneous WE-RE and RE-CE voltage and EIS techniques). Needs to be possible to self-calibrate. Software should be freely available and not require licenses or cloud computing. Needs to be able to interface with LabView virtual instrument software. needs to be able to load prior settings upon loading a previously-run data file. Should be able to group or synchronize channels for duplicate measurements. Software should be able to fit and simulate cyclic voltammetry curves. EIS analysis tools should include Mott-Schottky analysis. Should be able to run multiple sets of settings in a specific order utilizing a batch mode.
Attachments
Files listed on SAM.gov. The ones we hold a copy of download from here; the rest open SAM's download, which needs no account.
- Request for Proposal - 371738.pdf.pdf187 KB, April 29, 2025
- Attachment D-Bill of Materials.pdf76 KB, April 29, 2025
- Attachment C-Reps-and-Certs-Abbreviated - 02.10.2025.pdf1.1 MB, April 29, 2025
- Attachment B-Prime Supplemental Flowdown Document.pdf245 KB, April 29, 2025
- Attachment A-Terms and Conditions-Purchase Order Commercial items and Services Under Simplified Acquisition Threshold.pdf243 KB, April 29, 2025
Publications
Every notice SAM.gov issued under this solicitation number, oldest first. Each is a separate record on SAM.
April 29, 2025
Solicitation
Due May 7, 2025 at 5:00 PM EDT. SAM.gov, notice 5eb79ce717b44e8e81fc7ecbd1313665
1 revision on SAM
- April 29, 2025: publish, solicitation
Points of contact
- Morgan Halehalemt@ornl.gov6155199527