# SP-300 Potentiostat, Potentiostat/galvanostat board and a standard cell cable with electrometer

Canonical: https://sealedbidder.com/opportunities/rfq371738

- Solicitation number: RFQ371738
- Notice type: Solicitation
- Status: Closed. Deadline was May 7, 2025 at 5:00 PM EDT
- Department: Department of Energy
- Contracting office: ORNL Ut-Battelle LLC-DOE Contractor (899042)
- NAICS: 334516 Analytical Laboratory Instrument Manufacturing
- Product or service code: 6630 Chemical Analysis Instruments
- Small business size standard: 1,000 employees (13 CFR 121.201, NAICS 334516)
- Place of performance: Oak Ridge, Tennessee
- Who can bid, by location: Anywhere in the US. Location does not limit who can bid: a product shipped to the buyer, research at the performer's facility, or work at the contractor's site. A state shown is the delivery point.
- County: Anderson County (FIPS 47001). https://sealedbidder.com/counties/anderson-county-tn-47001
- City: Oak Ridge. https://sealedbidder.com/cities/oak-ridge-tn-4755120
- First posted: April 29, 2025
- Last posted: April 29, 2025
- SAM.gov: https://sam.gov/workspace/contract/opp/5eb79ce717b44e8e81fc7ecbd1313665/view

## Reading this notice

- Kind of notice: Solicitation. A formal request for proposals, quotes or bids. Offers are due by the response deadline.
- Who may compete: SAM.gov lists no set-aside on this notice, so it is not reserved for small businesses. Small, under NAICS 334516 (analytical laboratory instrument manufacturing), means no more than 1,000 employees, counted with affiliates.
- Time to respond: Bidders had 8 days from the day the solicitation was posted, April 29, 2025, to the deadline.

## Description

SP-300 Potentiostat, Potentiostat/galvanostat board and a standard cell cable with electrometer

Request for the exact or domestic equivalent multi-channel, modular design potentiostat/galvanostat/EIS analyzer based on a portable chassis is required. Needs to be capable of USB and ethernet communication. Needs the following operational modes: floating, grounded, counter electrode to ground, working electrode to ground. EIS experiment functionality from 10 micro-Hz to 7 MHz. No separate analyzer should be required. Both sine and multi-sine techniques need to be available. Current and voltage should be possible to be measured simultaneously, with compliance voltage of +/- 12 Volt and reference voltage adjustable between -10 to +10 volts, with same range between working and reference electrodes. Resolution must be programmable from 300 micro-volts down to 1 micro-volts. Needs to apply and measure currents from 1 pico-amp to 500 milli-amps, a current control resolution of 0.004% and accuracies to 0.1% of FSR. Should be possible to be interfaced with boosters of 1,2,4 or 10 Amps on the same chassis. Multi-amp boosters should be possible to be paired for higher currents. Should be expandable to include a linear scan generator for true analog waveforms, up to 1 MV/s with data aquisition rates as fast as 1 micro-second. Requires an ultra-low current option to +/- 100 femto-amps on 1 pico-amp range with greater than 10^14 ohms input impedance. Should be capable of switching control modes in under 5 microseconds, without adverse spiking of the current in the connected cell. Requires an external input for control of potential or current via external analog input from separate device. Needs to have at least nine user-selectable bandwidths for full stability control on channels. Analog filters (50 kHz, 1 kHz, 5 Hz) as well as digital filtering in software. Needs to be able to acquire data at 20 micro-second frequency. Needs to be capable of 2-5 terminal measurements (for simultaneous WE-RE and RE-CE voltage and EIS techniques). Needs to be possible to self-calibrate. Software should be freely available and not require licenses or cloud computing. Needs to be able to interface with LabView virtual instrument software. needs to be able to load prior settings upon loading a previously-run data file. Should be able to group or synchronize channels for duplicate measurements. Software should be able to fit and simulate cyclic voltammetry curves. EIS analysis tools should include Mott-Schottky analysis. Should be able to run multiple sets of settings in a specific order utilizing a batch mode.

## Attachments

- [Request for Proposal - 371738.pdf.pdf](https://sam.gov/api/prod/opps/v3/opportunities/resources/files/fb82c6d4f2ec4731a7a9e3556e6cb82e/download)
- [Attachment D-Bill of Materials.pdf](https://sam.gov/api/prod/opps/v3/opportunities/resources/files/376f29cc84614be8a6a1db2dcdb199fd/download)
- [Attachment C-Reps-and-Certs-Abbreviated - 02.10.2025.pdf](https://sam.gov/api/prod/opps/v3/opportunities/resources/files/4e47626b22674c0d9d39e5c8773d311f/download)
- [Attachment B-Prime Supplemental Flowdown Document.pdf](https://sam.gov/api/prod/opps/v3/opportunities/resources/files/79621d980da7455ab9a22d5cb1eda204/download)
- [Attachment A-Terms and Conditions-Purchase Order Commercial items and Services Under Simplified Acquisition Threshold.pdf](https://sam.gov/api/prod/opps/v3/opportunities/resources/files/c5aeb31e8822440799e722e7db0ab1c7/download)

## Open now in the same industry

59 open in NAICS 334516.

- Adjustable-Volume Laboratory Micropipettes and Pipettors (PR519028), due November 25, 2026 at 5:00 PM EST. https://sealedbidder.com/opportunities/pr519028
- ARBITRARY WAVEFORM GENERATOR (N0016427Q0014), due November 25, 2026 at 2:00 PM EST. https://sealedbidder.com/opportunities/n0016427q0014
- Brand Name or Equal Teledyne Instruments Liquid Sy (89243327QEX000002), due November 25, 2026 at 5:00 PM EST. https://sealedbidder.com/opportunities/89243327qex000002

## Publications

- April 29, 2025: Solicitation, due May 7, 2025 at 5:00 PM EDT. Notice 5eb79ce717b44e8e81fc7ecbd1313665. https://sam.gov/workspace/contract/opp/5eb79ce717b44e8e81fc7ecbd1313665/view

## Points of contact

- Morgan Hale, halemt@ornl.gov, 6155199527

---
Source: SAM.gov Contract Opportunities bulk extract. Confirm deadlines on SAM.gov before responding. Cite https://sealedbidder.com/opportunities/rfq371738.