Presolicitation, total small business set-aside
Water Cooled X-Ray System
A24-007
Department of the Air Force, FA8125 AFSC Pzima. Other Measuring and Controlling Device Manufacturing.
Response deadline
June 4, 2025 at 11:00 AM EDT
Closed 539 days ago. Posted May 29, 2025. Scheduled to archive June 19, 2025.
Get the next one like this from Department of the Air Force by email, the day it is posted.Find my opportunities
This one closed in 2025
49 like it are open now in other measuring and controlling device manufacturing
- HPLC Core System
Food Safety and Inspection Service, Usda, Fsis, Oas PCMB
PresolicitationNAICS 334519Coral Gables, FL1178021Due todayNov 25 - ADAPTER,SPEEDOMETER
Defense Logistics Agency, DLA Aviation
Combined synopsis and solicitationNAICS 334519Ship to the buyerSPE4A527T3404Due tomorrowNov 25 - ADAPTER,SPEEDOMETER
Defense Logistics Agency, DLA Aviation
Combined synopsis and solicitationNAICS 334519Ship to the buyerSPE4A527T3405Due tomorrowNov 25
All 49 open in other measuring and controlling device manufacturing
Reading this notice
Worked out from the publications and the award on this page.
- Kind of notice
- Presolicitation. Advance notice that a solicitation is coming, usually within a few weeks. Glossary
- Who may compete
- Total small business set-aside. Only small businesses may compete. Small, under NAICS 334519 (other measuring and controlling device manufacturing), means no more than 600 employees, counted with affiliates. Open Small business opportunities
- What comes next
- No solicitation was published under this number after the presolicitation of May 29, 2025. An agency sometimes issues the solicitation under a new number.
Size standard from 13 CFR 121.201. The solicitation itself is what binds: check the code and standard it states.
Description
As published on SAM.gov.
The Air Force intends to solicit the following supply:
76 AMXG is requesting utilization of funds for Request for the purchase of 4 ea.
Water Cooled Xray System to include the following:
Input Line Requirements:
- Automatically adapts to input line voltage100-130Vac, 50/60Hz, 20 Amperes maximum.
- 200-250Vac, 50/60Hz, 10 Amperes maximum
- 20 to 160kV, 0.5mA to 5.0mA (800 watts max.)
- Constant potential, end-grounded anode, liquid cooled version
X-Ray Tube Window:
- Beryllium 0.8mm (directional) 40° directional
Radiation Output:
- 14R/min at 50cm filtered with 0.5 inches (12.7mm)
- Aluminum at 160kV, 5mA
Radiation Leakage:
- Less than 2.0 Roentgens per hour at 1 meter from the X-Raytube target
- Effective Focal Spot: Standard: 0.7mm x 1.0mm (0.03in x 0.04in) EN125430.5mm x 0.5mm (0.02in x 0.02in) nominal IEC60336
Ambient Temp:
- 100% duty cycle @ 120°F (49°C)
- Duty Cycle: 100%
Storage Temp:
- -30°F to 160°F (-35°C to 71°C)
Anode Cooling:
- Liquid coolant solution closed loop between X-Ray tube anode and cooling unit.
- Liquid Cooling Unit Weight: 54lbs. (15.4kg) approx.
- Tube Head Dimensions: No greater than 7.25" Diameter, 30" Length
- Liquid Cooled: 29lbs. (13.15kg)
X-Ray Control Unit:
- X-Ray Control Unit Weight: 26lbs. (11.8kg) approx
Safety Devices:
- Tubehead Pressure Relief Valve
- Tubehead Thermal Cut-Out
- Tubehead Pressure Gauge
- Tubehead Low Pressure Cut-Out @ 25psi (1.72 Bar)
- Coolant Flow Sensor (liquid cooled only)
- Control Unit Safety Keyswitch
- Microcomputer-based Self Diagnostics
- Continuous Exposure Parameter Display
Standard Accessories:
- Operation manual
- Tubehead carrying case
- Tubehead Cable- 100’ with strain relief
- Extra key (1) for Control Unit Safety Lock
- Power cable - 25’ with strain relief
- Coolant hose - twin, 50’ with self-sealing terminations
- (liquid cooled only)
- Cooler power cable
- GUI Control Software for the GUI is specifically designed for controlling series systems. As an alternative to the front panel control, the GUI will allow the user to control all necessary functions of the system from a user-friendly windows-based menu. Additionally, the GUI can be used as a diagnostic tool when the system is controlled via the front panel.
A Firm Fixed Price type contract is contemplated.
Estimated Period of Performance: 360 days ARO
This is a small business set-aside
System for Award Management (SAM) is required to receive any award. Contractors can register electronically at https://www.sam.gov/portal/public/SAM/. The NAICS code for this requirement is 334519 with a size standard of 600 EMP.
Interested persons may identify their interest and capability to respond to the requirement by submitting a proposal. Electronic procedures will be used for this solicitation through Beta.sam.gov at https://sam.gov/content/home. No telephone requests. See Note 22. Authority is 10 USC 2304 (c)(1).
For further information contact:
Sondra Groth
Contracting Specialist
Sondra.groth@us.af.mil
Publications
Every notice SAM.gov issued under this solicitation number, oldest first. Each is a separate record on SAM.
May 29, 2025
Presolicitation
Due June 4, 2025 at 11:00 AM EDT. SAM.gov, notice baf436a0bbc54782ad278478ae94cd8a
2 revisions on SAM
- June 13, 2025: publish, solicitation
- May 29, 2025: publish, presolicitation
Points of contact
- Sondra GrothSondra.Groth@us.af.mil
- Sheridan Robisonsheridan.robison@us.af.mil