Combined synopsis and solicitation, total small business set-aside
Tokai Hit STXG-GSI2X stage top incubator for Leica Stellaris STED microscope
75N98026Q00794
National Institutes of Health, National Institutes of Health Olao. Analytical Laboratory Instrument Manufacturing.
Response deadline
July 25, 2026 at 9:00 AM EDT
Closed 124 days ago. Posted July 15, 2026. Scheduled to archive July 26, 2026.
Get the next one like this from National Institutes of Health by email, the day it is posted.Find my opportunities
This one closed in 2026
55 like it are open now in analytical laboratory instrument manufacturing
- Notice of Intent - Illumina MiSeq i100 series instruments
National Institutes of Health, National Institutes of Health - CC
PresolicitationNAICS 334516Ship to Bethesda, MDNOI-OLAO-CC-27-000836Due tomorrowNov 27 - Graphite Furnace Atomic Absorption Spectrometer
National Institute of Standards and Technology, Department of Commerce NIST
Sources soughtNAICS 334516Ship to Gaithersburg, MDNIST-MML-FY27-SS01Due in 2 daysNov 28 - Intent to Solicit Only One Source for Six (6) Twinleaf Pulsed Pump Magnetometer (part number: PPM-RS422) and one (1) Twinleaf 8 Channel Synchronization Hub (part number: HUB-RS422-8) manufactured by Twinleaf.
Department of the Army, W6QK ACC-APG Adelphi
SolicitationNAICS 334516Ship to MDW911QX27QA015Due in 5 daysDec 1
All 55 open in analytical laboratory instrument manufacturing
Reading this notice
Worked out from the publications and the award on this page.
- Kind of notice
- Combined synopsis and solicitation. A streamlined commercial-item buy where the notice itself is the solicitation. Glossary
- Who may compete
- Total small business set-aside. Only small businesses may compete. Small, under NAICS 334516 (analytical laboratory instrument manufacturing), means no more than 1,000 employees, counted with affiliates. Open Small business opportunities
- Time to respond
- Bidders had 10 days from the day the solicitation was posted, July 15, 2026, to the deadline.
Size standard from 13 CFR 121.201. The solicitation itself is what binds: check the code and standard it states.
Description
As published on SAM.gov.
Requirements:
Provides accurate dynamic control of temperature, humidity, CO2 and oxygen to maintain required cell culture conditions for live cell confocal and STED super-resolution imaging. The stage top incubator is required in order to perform live cell imaging with the unique resolution capabilities of our Leica Stellaris 5 STED microscope. It currently does not have a stage top incubator, and therefore, is unable to perform this type of imaging requested by NCI researchers.
Stage top incubator controller must have two feedback modes, chamber temperature or specimen temperature, while it also continuously monitors the room temperature to maintain specimen temperature within set point anywhere between 25 – 50 degrees Celsius. The built-in digital gas mixer controller accepts 100% CO2 from lab supply to regulate CO2 concentration at the sample, with a CO2 range of 5.0 to 20% with real-time sample feedback. In addition, humidity is actively controlled with an internal humidifier to sense sample chamber humidity. The electrically heated stage top incubator chamber must have magnets for easy interchange of specimen holders, including inserts for 35 mm coverglass dish, 50/60mm coverglass dish, chamber slide, coverglass chamber, and open frame insert for imaging multi-well chambers. A flexible heated collar that accommodates a range of objective lens diameters must also be included.
The components of the stage top incubator must include:
- Tokai Hit STXG-GSI2X-SET Stage Top Incubator for Super Z galvo stage (inverted microscope stands)
- Pure (100%CO2) gas use (110- 240V) for individual temperature settings to create optimum temperature and humidity conditions without difficulty for long term live cell imaging.
- Setting temperature range from 25 to 50°C.
- Dimensions of 112 x 87 mm compatible with Leica Super Z galvo stage on existing Leica Stellaris 5 STED microscope.
- the lightest chamber (less than 140g) for Super Z Glavo stage
- temperature controller with built-in digital gas mixer (CO2: 5.0 - 20.0%)
- Objective heater
- Magnetic Dish holder and fixing lid for 35mm, 50/60mm dish, chamber slide, slide glass and chamber slide use.
- External Sensor, enable to autoclave repeatability for Sample Feedback regulation.
- Tokai Hit original data logging, temp/CO2 program control, software compatible to use with LAS-X software
Attachments
Files listed on SAM.gov. The ones we hold a copy of download from here; the rest open SAM's download, which needs no account.
- Requirement 33.docx17 KB, July 15, 2026
Publications
Every notice SAM.gov issued under this solicitation number, oldest first. Each is a separate record on SAM.
July 15, 2026
Combined synopsis and solicitation
Due July 25, 2026 at 9:00 AM EDT. SAM.gov, notice 7f0309dced09447697a2aca2279013b4
1 revision on SAM
- July 15, 2026: publish, combined synopsis and solicitation
Points of contact
- John Burtonburtonjo@mail.nih.gov