Combined synopsis and solicitation, total small business set-aside
Optical Spectrum Analyzer
W9132V24S1479
Department of the Army, W2R2 USA Engr R & D CTR. Semiconductor and Related Device Manufacturing.
Response deadline
September 6, 2024 at 11:00 AM EDT
Closed 812 days ago. Posted August 30, 2024. Scheduled to archive September 21, 2024.
Get the next one like this from Department of the Army by email, the day it is posted.Find my opportunities
This one closed in 2024
37 like it are open now in semiconductor and related device manufacturing
- MICROCIRCUIT,HYBRID
Defense Logistics Agency, DLA Weapons Support
Combined synopsis and solicitationSmall businessNAICS 334413Ship to the buyerSPE7M527T2934Due tomorrowNov 27 - LIGHT EMITTING DIODE
Defense Logistics Agency, DLA Weapons Support
Combined synopsis and solicitationWOSBNAICS 334413Ship to the buyerSPE7M527U0078Due tomorrowNov 27 - LIGHT,INDICATOR
Defense Logistics Agency, DLA Troop Support
Combined synopsis and solicitationNAICS 334413Ship to the buyerSPE8E727T0533Due tomorrowNov 27
All 37 open in semiconductor and related device manufacturing
Reading this notice
Worked out from the publications and the award on this page.
- Kind of notice
- Combined synopsis and solicitation. A streamlined commercial-item buy where the notice itself is the solicitation. Glossary
- Who may compete
- Total small business set-aside. Only small businesses may compete. Small, under NAICS 334413 (semiconductor and related device manufacturing), means no more than 1,250 employees, counted with affiliates. Open Small business opportunities
- Time to respond
- Bidders had 7 days from the day the solicitation was posted, August 30, 2024, to the deadline.
Size standard from 13 CFR 121.201. The solicitation itself is what binds: check the code and standard it states.
Description
As published on SAM.gov.
The Geospatial Research laboratory needs an optical spectrum analyzer to be used in the experimental setup as part of ELS3D project. The GRL must procure this spectrum analyzer so the team at GRL can characterize laser beams in the photonics lab. Those lasers will be used as a component in the development of a high altitude lidar system. The spectrum analyzer will also be utilized in future 6.1 projects to measure the spectrum of gas particles, which means the resolution must be fine enough to do material characterization of the gas. The optical spectrum analyzer’s performance needs to meet certain specifications provided in the table below.
System Requirements (Table 1): Table 1, entitled "System Requirements," provides the key specifications we seek to satisfy.
Selection Process: Technical experts at GRL will review each quote offered to GRL in response to this solicitation. To be considered, your quote must show that your product will meet and deliver all the specifications in Table 1. The vendor is requested to provide clear quotes showing that their products meet all the listed requirements to ensure the technical experts see that an offered product will meet all the required standards. Quotes that are ambiguous or are missing required information may be disqualified.
Selection will be based on the lowest price, which is technically acceptable.
Table 1. System Requirements Specifications Performance Requirement Spectral Resolution <= 2 GHz Spectral Accuracy +/- 2 ppm Level Sensitivity -60 dBm Wavelength meter Accuracy +/- 1 ppm Power Level Accuracy +1 dB Optical Rejection Ratio 30 dB Optical Input Fiber coupled port and free-space port required Hose for optical purging Required Polarization Dependence +/- 1 dB GUI and analysis Tools Operates device and includes spectroscopic analysis Weight <60 lbs Alignment laser
Publications
Every notice SAM.gov issued under this solicitation number, oldest first. Each is a separate record on SAM.
August 30, 2024
Combined synopsis and solicitation
Due September 6, 2024 at 11:00 AM EDT. SAM.gov, notice 290c27848f5e42d996258eaea3063801
Points of contact
- JESSICA WILSONJESSICA.L.WILSON2@USACE.ARMY.MIL