# Automated Media Preparation Instrument for Dissolution Testing

Canonical: https://sealedbidder.com/opportunities/ssnoc2026133856

- Solicitation number: SSN_OC_2026_133856
- Notice type: Sources sought
- Status: Closed. Deadline was May 19, 2026 at 12:00 PM EDT
- Department: Department of Health and Human Services
- Agency: Food and Drug Administration
- Contracting office: FDA Office of Acq Grant SVCS (75F401)
- NAICS: 334516 Analytical Laboratory Instrument Manufacturing
- Product or service code: 6632 Analysis Instruments, Medical Laboratory Use
- Set-aside: Total small business set-aside
- Small business size standard: 1,000 employees (13 CFR 121.201, NAICS 334516)
- Place of performance: Irvine, California
- Who can bid, by location: Anywhere in the US. Location does not limit who can bid: a product shipped to the buyer, research at the performer's facility, or work at the contractor's site. A state shown is the delivery point.
- County: Orange County (FIPS 06059). https://sealedbidder.com/counties/orange-county-ca-06059
- City: Irvine. https://sealedbidder.com/cities/irvine-ca-0636770
- First posted: May 13, 2026
- Last posted: May 13, 2026
- SAM.gov: https://sam.gov/workspace/contract/opp/d0de35874432456e9b046fc3f98bc6e5/view

## Reading this notice

- Kind of notice: Sources sought. Market research. The agency wants to know who could do the work before it decides how to compete it.
- Who may compete: Total small business set-aside. Only small businesses may compete. Small, under NAICS 334516 (analytical laboratory instrument manufacturing), means no more than 1,000 employees, counted with affiliates.
- What comes next: No solicitation was published under this number after the sources sought of May 13, 2026. An agency sometimes issues the solicitation under a new number.

## Description

This requirement is being acquired as a brand-name-or-equal procurement. Equal products meeting all stated salient characteristics will be considered.

		The instrument shall be capable of delivering a minimum volume of 250ml and a maximum volume of at least 1000ml.

		The instrument shall have a dispensing accuracy of &plusmn;5mL or less.

		The instrument shall be equipped with in-line heating that is capable of heating dissolution media to at least 45&deg;C.

		The instrument shall have a temperature accuracy of &plusmn;2&deg;C or better.

		The instrument shall be capable of degassing media up to a maximum of 3.0ppm.

		The instrument shall be compatible with media containing up to a maximum of 2% surfactants.

		The instrument shall have an onboard controller and user interface to select stored programs and set parameters for manual use.

		The instrument shall be capable of storing programmed methods and running in automatic or manual mode when preparing dissolution media.

		The instrument shall be capable of storing dispensing reports that include the volume and temperature dispensed.

		The instrument shall be portable with a cart or other mechanism, included, to allow the instrument to be moved to each dissolution instrument for use.

		The instrument shall be equipped with a dispensing nozzle that allows for dissolution vessels to be filled in place, while on the dissolution apparatus.

		Electrical requirement must be compatible with a standard US outlet, 115V, 50 or 60 Hz.

Safety Requirements

The Contractor shall ensure all equipment complies with applicable laboratory safety standards and manufacturer safety requirements. All installation activities shall be performed in accordance with applicable safety regulations to prevent hazards to personnel and facilities.

Trade and Service Specifications

	The instrument must be a newly manufactured unit, not used and refurbished or previously used for demonstration.

	FOB Point destination to include inside delivery and clean-up of area after installation.

	The entire system must be warranted for parts and labor for a minimum of 12 calendar months from the date of formal government acceptance.  The vendor must also be capable of servicing the instrument through the covered warranty period.  The system must include, at a minimum: coverage on all non-consumable items and parts supplied including base instrument, factory-certified replacement parts, engineer labor and travel costs. Any equipment repair and maintenance work shall be performed by an OEM-trained engineer.  This factory-trained engineer shall have (verified by the OEM) the following: 1) access to OEM factory telephone support; 2) access to the most current OEM factory training for both hardware and software components; and 3) access to all current OEM factory parts, not build-to-order parts.  The OEM-trained service engineer shall not use salvaged parts from other instruments for performing maintenance and repairs.  All parts used in PM and repairs must be guaranteed, factory-tested, OEM quality parts.

	Instrument operators shall have access, for the duration of the 12-month warranty period, to a technical representative call center at no additional charge, for technical assistance and trouble-shooting, which is staffed by senior engineers to provide a high level of expertise for troubleshooting the instrument.

	Operator level training for 2 users at the delivery onsite location shall be provided at time of installation or shortly thereafter (within 30 days), such that operators are able to independently operate the instrument and included software to conduct routine analyses on the instrument.

	The Contractor shall provide installation by a trained and certified field service engineer who shall perform the necessary verification testing to demonstrate that the instrument meets the Original Equipment Manufacturer&rsquo;s operational standards following installation. Consumables used for the installation shall be included.

Records and Reports

The Contractor shall, commensurate with the completion of each service call (inclusive of warranty service), provide the end-user of the equipment with a copy of a field service report/ticket identifying the equipment name, manufacturer, model number, and serial number of the equipment being serviced/repaired and detailing the reason for the service call, a detailed description of the work performed, the test instruments or other equipment used to affect the repair or otherwise perform the service, the name(s) and contact information of the technician who performed the repair/service, and for information purposes, the on-site hours expended and parts/components replaced.  In addition, the Contractor shall provide monthly reports to the FDA Project Officer and Contract Specialist, not later that the 5th workday following the end of each month, summarizing all maintenance and repair activities (including warranty work) for the previous month (during months that work is performed).

## Attachments

- [SSN.docx](https://sam.gov/api/prod/opps/v3/opportunities/resources/files/b6b973ff0865440686d390307497185c/download)

## Open now in the same industry

53 open in NAICS 334516.

- ChemiDoc Go Imaging System (75N98026Q00083), due November 28, 2026 at 9:00 AM EST. https://sealedbidder.com/opportunities/75n98026q00083
- Spectrum Analyzer (N6426727Q4009), due December 2, 2026 at 3:00 AM EST. https://sealedbidder.com/opportunities/n6426727q4009
- SPEAR Beam Position Monitors (BPMs) (SLAC_RFP369851_(AT)), due December 4, 2026 at 8:00 PM EST. https://sealedbidder.com/opportunities/slacrfp369851at

## Publications

- May 13, 2026: Sources sought, due May 19, 2026 at 12:00 PM EDT. Notice d0de35874432456e9b046fc3f98bc6e5. https://sam.gov/workspace/contract/opp/d0de35874432456e9b046fc3f98bc6e5/view

## Points of contact

- Reagan Davis, reagan.davis@fda.hhs.gov, 3017960910

---
Source: SAM.gov Contract Opportunities bulk extract. Confirm deadlines on SAM.gov before responding. Cite https://sealedbidder.com/opportunities/ssnoc2026133856.