# CHIPS R&D: Fourier-Transform Infrared (FTIR) Spectrometer – Combined Sources Sought/Notice of Intent to Sole Source

Canonical: https://sealedbidder.com/opportunities/nistss26chips60

- Solicitation number: NIST-SS26-CHIPS-60
- Notice type: Special notice
- Status: Closed. Deadline was January 26, 2026 at 11:00 AM EST
- Department: Department of Commerce
- Agency: National Institute of Standards and Technology
- Contracting office: Department of Commerce NIST (1333ND)
- NAICS: 334519 Other Measuring and Controlling Device Manufacturing
- Product or service code: 6640 Laboratory Equipment and Supplies
- Small business size standard: 600 employees (13 CFR 121.201, NAICS 334519)
- Place of performance: Gaithersburg, Maryland
- Who can bid, by location: Anywhere in the US. Location does not limit who can bid: a product shipped to the buyer, research at the performer's facility, or work at the contractor's site. A state shown is the delivery point.
- County: Montgomery County (FIPS 24031). https://sealedbidder.com/counties/montgomery-county-md-24031
- City: Gaithersburg. https://sealedbidder.com/cities/gaithersburg-md-2431175
- First posted: January 12, 2026
- Last posted: January 14, 2026
- SAM.gov: https://sam.gov/workspace/contract/opp/985467fb22544c1f87f13db62f782781/view

## Reading this notice

- Kind of notice: Special notice. Industry days, draft requirements, program announcements and other notices that are not a solicitation.

## Description

***THIS IS A COMBINED SOURCES SOUGHT NOTICE AND NOTICE OF INTENT TO SOLE SOURCE***

The National Institute of Standards and Technology (NIST) Chemical Process and Nuclear Measurements Group is seeking a research-grade Fourier Transform Infrared (FT-IR) Spectrometer. The Group is involved in a CHIPS R&D Metrology project to characterize the stability of biofunctional semiconductor interfaces in chip-based biosensors with a focus on improving manufacturability. As a part of this program, NIST aims to develop operando methods for biointerface characterization using optical, electrical, and electrochemical measurements on surfaces functionalized with self-assembled monolayers (SAMs) and bioreceptors. The Chemical Process and Nuclear Measurements Group in the Chemical Science Division has specialized expertise in developing in situ and operando infrared measurements on various chemical systems. In support of this work, the group requires a research-grade Fourier Transform Infrared (FT-IR) Spectrometer.

The CHIPS Metrology program develops and advances cutting edge metrology capabilities for members of the US semiconductor manufacturing ecosystem.  This NIST conducted research program works with device manufacturers, tool vendors, materials suppliers, and other organizations to address critical metrology gaps to spur innovation within seven grand challenge areas.  For more information on CHIPS Metrology, please visit https://www.nist.gov/chips/research-development-programs/metrology-program.

The Chemical Process and Nuclear Measurements Group is seeking a research-grade Fourier Transform Infrared (FT-IR) Spectrometer, which will be used to measure the stability and failure mechanisms of chip-based biosensors. The measurements are a critical component of a CHIPS R&D Metrology program to develop a relationship between functionalization, performance, and lifetime of biosensors built using semiconductor technology. Understanding the specific mechanisms of interface fabrication and degradation will accelerating biosensor development and improve manufacturing processes critical for commercial viability.

NIST is seeking information from sources that may be capable of providing a solution that will achieve the objectives described above, in addition to the following essential requirements:

Line Item 0001: FT-IR Spectrometer

Description: Infrared spectrometer configured with KBr beamsplitter and windows, DLaTGS detector, fast scan option, external detector positions, trigger electronics, time-resolved measurement capability, and data processing software

Quantity:  1

	Technical Specifications for Hardware

		Spectral range: 8,000 cm-1 to 350 cm-1

		Spectral range extension: Field-upgradable to 15 cm-1 to 28,000 cm-1

		Spectral resolution: 0.16 cm-1 or better

		Wavenumber accuracy: 0.005 cm-1 or better at 1,554 cm-1

		Photometric accuracy: 0.1% transmission or better

		Signal-to-noise ratio: 10,000:1 or better at 2,000 cm-1, 4 cm-1 resolution, 5 s acquisition time (sample and reference channels)

		Interferometer type: Permanently aligned interferometer with no adjustments for maximal long-term stability

		Interferometer speed: 2 kHz to 160 kHz, or faster

		Laser: Stabilized HeNe

		Scan type: Rapid scan (continuous scan)

		Spectral acquisition rate: 70 spectra/s or greater at 16 cm-1 resolution

		Analog-to-digital converter (ADC):

			Dual-channel, 24-bit

			On-detector pre-amp and ADC

		Optics bench purging:

			Built-in desiccant cartridges

			Dry air or nitrogen purging from external supply

		Light source: Stabilized, air-cooled, globar

		Beamsplitter: Broadband KBr, 10,000 cm-1 to 400 cm-1

		Detector:

			Temperature-controlled DTGS installed in internal compartment

			Additional internal detector position for one MCT or InSb detector

			Software-controlled motorized beam steering to each detector position

		Beam outlet ports: Two outlets

			First outlet on the right side of the spectrometer bench

			Second outlet on the left side of the spectrometer bench

			Software-controlled motorized beam steering for each output

		Windows:

			Sample compartment: KBr

			Beam outlets (2): KBr

		Aperture wheel (J-stop): 250 um to 8 mm in diameter, at least 10 positions

		Filter wheel:

			Compatible with 25 mm filters, at least 8 position

			Installed filters: NG-4 glass, NG-9 glass, NG-11 glass, polystyrene

		External input-output: At least 4 digital I/O channels to send or receive TTL signals

			Hardware interface: BNC connectors

		External detector interface: At least 2 external detector channels

		Compatibility with Bruker optics and accessory plates:

			Compatible with Bruker beamsplitters from Invenio R

			Compatible with Bruker sources from Invenio R

			Compatible with Bruker detectors from Vertex 70 and Invenio R

			Compatible with Bruker QuickLock baseplates for accessories

		Communications interface: Ethernet

	Technical Specifications for Software

		Scope: Single software package for data acquisition, automation, and analysis

		Operating system compatibility: Windows 11 64-bit

		Installation: User-installable on government owned PC

		Data acquisition:

			Time-resolved rapid-scan mode with double-sided forward/backward interferogram processing (split interferograms)

			Repetitive rapid-scan measurements initiated by external TTL signals

			Sequence-based data acquisition, including nested loops for spectral co-adding over multiple iterated acquisitions

				Control over digital IO lines during a sequence

				Ability to add arbitrary time delays to a sequence

				Ability to wait for multiple trigger signals at different points during a sequence

		Data storage: Single data file for storing single channel, interferogram, and absorbance/transmittance data with traceable log of all data transformations for both single spectra and time-series spectra

		Data processing:

			Ability to set FT parameters such as phase resolution, phase correction, apodization, and zero filling

			Ability to reprocess single channel data

			Ability to perform arbitrary spectral calculations such as addition, subtraction, averaging, and ratioing

			Ability to perform peak fitting and integration with adjustable peak models and background subtraction

		Data visualization:

			2D and 3D plotting of single spectra and time-series spectra

			Arbitrary scaling and manipulation of plotted spectra

			Stacked, tiled, and cascaded display of multiple single spectra

			Contour and surface plots of time-series spectra

		Time-resolved data processing:

			Ability to co-add and extract spectra from rapid-scan data

			Ability to reprocess rapid-scan interferograms using arbitrary reference spectra as background

		Data export: Batch exporting of multiple data files to human-readable text format such as CSV

		Instrument control: Graphical representation of beam path in spectrometer bench with software-controlled mirrors for path selection

		Optical component recognition: Automated recognition and configuration of installed optical components (sources, beamsplitter, detectors, accessories)

		Saved instrument configuration:

			Ability to save and recall the complete hardware configuration and acquisition parameters used for a given measurement using a single file

			Ability to save and recall rapid-scan automated sequences using a single file

		Programming interfaces:

			LabVIEW-compatible programming interface to execute measurements, acquire spectra, and read data files for post-processing

			Python programming interface to execute measurements, acquire spectra, and read data files for post-processing; ability to execute Python code in the FT-IR software

Line Item 0002:  InSb Detector

Description: Cryogenic InSb detector

Quantity: 1

	Technical Specifications

		Liquid-nitrogen cooled InSb detector

			10,000 to 1,850 cm-1 spectral range

			D* = 1.5x1011 cmHz1/2/W or better

			On-detector pre-amplifier and 24-bit ADC

		Mountable inside spectrometer bench

		Interface cable for connecting externally mounted detector to spectrometer bench

		External mounting hardware, such as dovetail guide to mount detector to custom optical apparatus

Line Item 0003:  MCT Detector

Description: Cryogenic MCT detector

Quantity: 1

	Technical Specifications

		Liquid-nitrogen cooled MCT detector

			12,000 to 600 cm-1 spectral range

			D* = 2x1010 cmHz1/2/W or better

			On-detector pre-amplifier and 24-bit ADC

		Mountable inside spectrometer bench

		Interface cable for connecting externally mounted detector to spectrometer bench

		External mounting hardware, such as dovetail guide to mount detector to custom optical apparatus

Line Item 0004:  Wafer ATR Accessory

Description: Prism coupled wafer ATR accessory

Quantity: 1

	Technical Specifications

		For attenuated total reflection (ATR) analysis of impurities in silicon wafers

		Silicon prism based unit for non-destructive mechanical coupling to wafers

		40 mm spacing between points coupling the wafer to the prism

		Silicon wafer samples act as internal reflection elements

		Sufficiently larger number of reflections for analysis of organic monolayers

		Mountable inside purged spectrometer bench

		Compatible with Bruker QuickLock baseplates

NIST conducted market research from November 2025 through November 2025 by speaking with colleagues, performing internet searches, and speaking with vendors to determine what sources could meet NIST&rsquo;s minimum requirements.  The results of that market research revealed that only Bruker Scientific LLC (UEI: MG2JF771VWU5) appears to be capable of meeting NIST&rsquo;s requirements.

HOW TO RESPOND TO THIS NOTICE

In responding to this notice, please DO NOT PROVIDE PROPRIETARY INFORMATION. Please include only the following information, readable in either Microsoft Word 365, Microsoft Excel 365, or .pdf format, in the response.  Submit the response by email to the Primary Point of Contact and, if specified, to the Secondary Point of Contact listed at the bottom of this notice as soon as possible, and preferably before the closing date and time of this notice.

	Provide the complete name of your company, address, name of contact for follow-up questions, their email, their phone number and, if your company has an active registration in https://sam.gov, your company&rsquo;s Unique Entity ID (UEI).

	Details about what your company is capable of providing that meets or exceeds NIST&rsquo;s minimum requirements.

	Whether your company is an authorized reseller of the product or service being cited and evidence of such authorization.

	Identify any aspects of the description of the requirements in the BACKGROUND section above that could be viewed as unduly restrictive or create unnecessary barriers that adversely affect your firm&rsquo;s ability to fully participate in a procurement for such services and explain why.  Please offer suggestions for how the requirements could be organized or structured to encourage the participation of small businesses.

	For the NAICS code

	Indicate whether your company is (a) a small business or (b) other than small business.  See the Table of Small Business Size Standards and the associated .pdf download file for small business size standards and additional information.

	If you believe the NAICS code listed in this notice is not the best NAICS code for the type of product addressed in this notice, identify an alternative NAICS code that you believe would be more appropriate for the planned procurement.

	If your firm has existing Federal Supply Schedule contract(s) or other contracts for products or services against which the Department may be able to place orders, identify the contract number(s) and other relevant information.

	Describe your firm&rsquo;s experience (as a prime, subcontractor, or consultant) providing the products or services described in Background section.

	Provide any other information that you believe would be valuable for the Government to know as part of its market research for this requirement.

	Please let us know if you would like to engage to get a better understanding of the requirement or need additional information about the Government&rsquo;s requirement for the products or services described in the Background section.

QUESTIONS REGARDING THIS NOTICE

Questions regarding this notice may be submitted via email to the Primary Point of Contact and the Secondary Point of Contact listed in this notice. Questions should be submitted so that they are received with 5 days of this posting.  Questions will be anonymized and answered via sources sought notice amendment following the question submission deadline.

IMPORTANT NOTES

The information received in response to this notice will be reviewed and considered so that the NIST may appropriately solicit for its requirements in the near future.

This notice should not be construed as a commitment by the NIST to issue a solicitation or ultimately award a contract.

This notice is not a request for a quotation. Responses will not be considered as proposals or quotations.

No award will be made as a result of this notice.

NIST is not responsible for any costs incurred by the respondents to this notice.

NIST reserves the right to use information provided by respondents for any purpose deemed necessary and appropriate.

ThaNk you for taking the time to submit a response to this request!

## Open now in the same industry

47 open in NAICS 334519.

- Request for Information (RFI): Vehicle Integrated Point Enhanced RADIAC (VIPER) - Tracked Ground Vehicles (VIPER_27-TRACK), due December 10, 2026 at 12:00 PM EST. https://sealedbidder.com/opportunities/viper27track
- REFERENCE STANDARD,PHY (SPE4A627U0220), due November 27, 2026. https://sealedbidder.com/opportunities/spe4a627u0220
- STARTER,MOTOR (SPE7L327T2652), due November 27, 2026. https://sealedbidder.com/opportunities/spe7l327t2652

## Publications

- January 12, 2026: Special notice, due January 26, 2025 at 11:00 AM EST. Notice a91ec32e59f74c9fb67209ad51e94332. https://sam.gov/workspace/contract/opp/a91ec32e59f74c9fb67209ad51e94332/view
- January 14, 2026: Special notice, due January 26, 2026 at 11:00 AM EST. Notice 985467fb22544c1f87f13db62f782781. https://sam.gov/workspace/contract/opp/985467fb22544c1f87f13db62f782781/view

## Points of contact

- Tracy Retterer, Tracy.retterer@nist.gov
- Forest Crumpler, forest.crumpler@nist.gov

---
Source: SAM.gov Contract Opportunities bulk extract. Confirm deadlines on SAM.gov before responding. Cite https://sealedbidder.com/opportunities/nistss26chips60.