# Ultra High-Performance Liquid Chromatography System

Canonical: https://sealedbidder.com/opportunities/nistmml25ssnb6450802500715

- Solicitation number: NISTMML25SSNB6450802500715
- Notice type: Sources sought
- Status: Closed. Deadline was April 7, 2025 at 2:00 PM EDT
- Department: Department of Commerce
- Agency: National Institute of Standards and Technology
- Contracting office: Department of Commerce NIST (1333ND)
- NAICS: 334516 Analytical Laboratory Instrument Manufacturing
- Small business size standard: 1,000 employees (13 CFR 121.201, NAICS 334516)
- Place of performance: Gaithersburg, Maryland
- County: Montgomery County (FIPS 24031). https://sealedbidder.com/counties/montgomery-county-md-24031
- City: Gaithersburg. https://sealedbidder.com/cities/gaithersburg-md-2431175
- First posted: March 6, 2025
- Last posted: March 6, 2025
- SAM.gov: https://sam.gov/workspace/contract/opp/115b2d2dd3904f889ce103f3ddf5a03a/view

## Reading this notice

- Kind of notice: Sources sought. Market research. The agency wants to know who could do the work before it decides how to compete it.
- Who may compete: SAM.gov lists no set-aside on this notice, so it is not reserved for small businesses. Small, under NAICS 334516 (analytical laboratory instrument manufacturing), means no more than 1,000 employees, counted with affiliates.
- What comes next: No solicitation was published under this number after the sources sought of March 6, 2025. An agency sometimes issues the solicitation under a new number.

## Description

This is a Sources Sought Notice ONLY.  Requests for copies of a solicitation will not receive a response.

This Notice is for planning purposes only and is not a Request for Proposal or Request for Quotation or an obligation on the part of the National Institute of Standards and Technology (NIST) for conducting a follow-on acquisition. NIST does not intend to award a contract on the basis of this notice or otherwise pay for the information requested. No entitlement or payment of direct or indirect costs or charges by NIST will arise as a result of submission of responses to this Notice and NIST use of such information. NIST recognizes that proprietary components, interfaces and equipment, and clearly mark restricted or proprietary components, interfaces and equipment, and clearly mark restricted or proprietary data and present it as an addendum to the non-restricted/non-proprietary information. In the absence of such identification, NIST will assume to have unlimited rights to all technical data in the information paper.

NO SOLICITATION DOCUMENTS EXIST AT THIS TIME.

Requirement: Ultra-Performance Liquid Chromatography System with Triple Quadrupole Mass Spectrometer

NIST is seeking information from sources with the following capabilities:

All of the following instruments must be new, not used or remanufactured equipment.

Triple Quadrupole Mass Spectrometer

Specifications:

	Capable of collision-induced dissociation (CID) using high-purity nitrogen as the collision gas

	Electrospray ion source with thermal gradient focusing technology capable of generating multiply-charged ions with heated nitrogen nebulizing gas capable of stable spray operation with liquid flow rates between 50 &micro;L/min and 2.5 mL/min

	Ion transfer capillary between the electrospray source and the entrance ion optics that has multiple channels for ion transmission to ensure measurement robustness and signal stability

	Software control of all operating gas flow rates and voltages

	Upgradable to perform atmospheric pressure chemical ionization (APCI)

	First and third quadrupoles capable of scanning between m/z 5 and m/z 3000

	Capable of using an integrated tuning and calibration solution that is completely software-controlled for remote operation capabilities

	Tuning and calibration automated through software control

	Includes a vacuum manifold that isolates the ion optics between the source and first quadrupole and allows the ion optics to be cleaned without completely venting the vacuum atmosphere of the mass spectrometer

	Software capable of molecule-specific tuning and signal optimization

	Quadrupole scanning speeds of &ge; 15,000 Da/s

	Multiple-reaction monitoring with a dwell time &le; 0.5 ms

	Capable of performing &ge; 500 multiple-reaction monitoring transitions per second

	First and third quadrupoles capable of scanning with a mass resolution of &le; 0.4 Da

	Capable of performing repetitive (N &ge; 5) peak area ratio determinations for LC-MS/MS analysis of stable isotope-labeled small molecules or peptide molecules and their natural, unlabeled form with an uncertainty for the peak area ratio of &le; 1 %

	Rough pumps for the mass spectrometer include sound-dampening covers

Complete ultra-performance liquid chromatography (UPLC) system with binary pump

Specifications:

	Consists of a binary pump with degassing capabilities, an autosampler, and a temperature-controlled column compartment

	Includes a stackable, modular UHPLC system

	Be able to be controlled from the same software that operates the mass spectrometer

UPLC Binary Pump

Specifications:

	Capable of producing binary solvent gradients

	Able to incorporate selection valves for each of the binary solvent channels that allow software selection of two or more solvents in the binary gradient production

	Minimum pumping flow rate must be &le; 0.001 mL/min; the maximum pumping flow rate must be &ge; 2.5 mL/min

	Pump flow rare accuracy must be &le; &plusmn; 1 %

	Capable of pumping to a backpressure of &ge; 1300 bar

	Software-controlled purge valve to facilitate remote pump operation

	Capable of pumping with pressure ripples &le; 1 % of total pressure

	Gradient mixer with a delay volume of &le; 50 &micro;L

	Integrated seal washing

	Integrated solvent degasser

	Integrated leak detection capabilities

	Automatic pumping compensation of solvent compressibility

	Pump resolution size must be &le; 500 pL

UPLC Autosampler

Specifications:

	Capable of injecting from 2-mL vials and 96-well plates

	Injection volumes between 0.1 &micro;L and 25 &micro;L

	Heating and cooling of the samples in a temperature range from 5 ˚C to 40 ˚C

	Programmable injection needle washing, both the internal and external parts of the injection needle, with &ge; 2 different wash solutions

	Operations at pressures of a pressure of &ge; 1300 bar

	Integrated leak detection

UPLC Temperature-controlled Column Compartment

Specifications:

	Cooling and heating column from &ge; 20 ˚C below room ambient and &ge; 100 ˚C

	Space for, at least, 2 standard length (150 mm) chromatography columns

	Integrated internal switching valve for multi-column applications

	Leak-free operation at a pressure of &ge; 1300 bar

	Integrated pre-column heat exchanger for solvents

	Integrated leak detection

Instrument Control and Data Analysis Software and Computers

Specifications:

	Instrument control software controls all operation functionality of the mass spectrometer and the liquid chromatography system

	Software to analyze LC-MS/MS data for both qualitative and quantitative analyses

	Software to analyze complex proteomic data that allows for database searching and deconvolution of spectra from multiply-charged ions

	Desktop computer running Windows 11 and the necessary peripheral devices such as networking cards, monitor, keyboard, and mouse

LC Bench

Specifications:

	Moveable instrument cart that can support all components of the UHPLC system and incorporates a power strip and solvent waste management.

MS Bench

Specifications:

	Moveable instrument cart that incorporates a power strip for the electrical connections of the mass spectrometer

	Supports an instrument weight of &ge; 300 lbs

Interested parties shall describe the capabilities of their organization as it relates to the equipment described above. NIST anticipates issuing a Request for Quotation in the second or third quarter of FY2025 and awarding a contract no later than the fourth quarter of FY2025.

NIST is seeking responses from all responsible sources, including large and small businesses. The small business size standard associated with the NAICS code for this effort, 334516, is 1000 employees. Please include your company&rsquo;s size classification and socio-economic status in any response to this notice.

After results of this market research are obtained and analyzed, NIST may conduct a competitive procurement and subsequently award a contract. Companies that can provide such equipment are requested to email a written response describing their abilities to jennifer.priebe@nist.gov no later than the response date for this sources sought notice.

The following information is requested to be provided as part of the response to this sources sought notice:

	Name, Address, Unique Entity Identifier (UEI), CAGE code, and point of contact information of your company.

	Any information on the company&rsquo;s small business certifications, if applicable.

	Description of your company&rsquo;s capabilities as they relate to the equipment described in this notice.

	A description of your company&rsquo;s previous experience providing the equipment described in this notice.

	Indication of whether the equipment described in this notice are currently offered via your company&rsquo;s GSA Federal Supply Schedule (FSS) contracts, Government-wide Acquisition Contracts (GWACs), or other existing Government-wide contract vehicles; and, if so, the contract number(s) for those vehicles.

	Any other relevant information that is not listed above which the Government should consider in finalizing its market research.

	Place of Manufacture of the equipment.

Responses are limited to a total of twelve (12) pages. The responses must be in MS Word format.  Pages shall be 8&frac12;-inch x 11-inch, using Times New Roman 11 Point Font. Each page shall have adequate margins on each side (at least one inch) of the page. Header/footer information (which does not include any information to be analyzed) may be included in the 1" margin space.

## Open now in the same industry

51 open in NAICS 334516.

- Graphite Furnace Atomic Absorption Spectrometer (NIST-MML-FY27-SS01), due November 28, 2026 at 2:00 PM EST. https://sealedbidder.com/opportunities/nistmmlfy27ss01
- Intent to Solicit Only One Source for Six (6) Twinleaf Pulsed Pump Magnetometer (part number: PPM-RS422) and one (1) Twinleaf 8 Channel Synchronization Hub (part number: HUB-RS422-8) manufactured by Twinleaf. (W911QX27QA015), due December 1, 2026 at 11:59 AM EST. https://sealedbidder.com/opportunities/w911qx27qa015
- Leica Microsystems, Inc. – Tokai Hit stage top incubator and Super Z Galvo stage (75N98027Q00115), due December 1, 2026 at 12:00 PM EST. https://sealedbidder.com/opportunities/75n98027q00115

## Publications

- March 6, 2025: Sources sought, due April 7, 2025 at 2:00 PM EDT. Notice 115b2d2dd3904f889ce103f3ddf5a03a. https://sam.gov/workspace/contract/opp/115b2d2dd3904f889ce103f3ddf5a03a/view

## Points of contact

- Jennifer Priebe, jennifer.priebe@nist.gov

---
Source: SAM.gov Contract Opportunities bulk extract. Confirm deadlines on SAM.gov before responding. Cite https://sealedbidder.com/opportunities/nistmml25ssnb6450802500715.