Special notice
Thermo-Optical Plane Source Thermal Conductivity Measurement Instrument
NB643090-25-01807
National Institute of Standards and Technology, Department of Commerce NIST. Analytical Laboratory Instrument Manufacturing.
Response deadline
July 7, 2025 at 1:00 PM EDT
Closed 507 days ago. Posted June 20, 2025. Scheduled to archive July 22, 2025.
Get the next one like this from National Institute of Standards and Technology by email, the day it is posted.Find my opportunities
This one closed in 2025
53 like it are open now in analytical laboratory instrument manufacturing
- PROBE,EDDY CURRENT
Defense Logistics Agency, DLA Aviation
Combined synopsis and solicitationSmall businessNAICS 334516Ship to the buyerSPE4A627T115JDue in 2 daysNov 27 - NCSS - Coldhead Exchange Parts and Calibration for
National Institutes of Health, National Institutes of Health Olao
Combined synopsis and solicitationNAICS 334516Ship to the buyer75N98027Q00098Due in 4 daysNov 30 - NOI to Sole Source ESLfemto Laser
Departmental Offices, Ibc Acq SVCS Directorate
PresolicitationNAICS 334516Ship to the buyer140D0426Q0057Due in 5 daysDec 1
All 53 open in analytical laboratory instrument manufacturing
Reading this notice
Worked out from the publications and the award on this page.
- Kind of notice
- Special notice. Industry days, draft requirements, program announcements and other notices that are not a solicitation. Glossary
Description
As published on SAM.gov.
FBO ANNOUNCEMENT: PRESOLICITATION NOTICE, NOTICE OF INTENT
ACTION CODE: SPECIAL NOTICE
SUBJECT: Sole source – Thermo-Optical Plane Source Thermal Conductivity Measurement Instrument
SOLICITATION NUMBER: NB643090-25-01807
RESPONSE DATE: 07/07/25
DESCRIPTION:
The National Institute for Standards and Technology (NIST), requires an instrument that can provide non-destructive, non-contact, and spatially resolved thermal conductivity measurements for solids, liquids, pastes, gels, and foams that have a broad range of shape, size, and surface roughness. NIST requires a direct measurement of thermal conductivity that does not require numerical input values for heat capacity, density, or sample dimensions, enabling high-throughput, rapid characterization and analysis. The instrument will generate reliable data for NIST researchers and for semiconductor industry partners to improve thermal management in semiconductor microelectronics packaging. To further this mission, NIST requires a Thermo-Optical Plane Source.
The Thermo-Optical Plane Source (TOPS) technique utilizes laser heating and infrared pyrometry to provide non-contact, non-destructive, and direct measurements of thermal conductivity. It can be implemented to characterize a wide variety of material forms, including ceramics, metals, pastes, liquids, polymers, composites, thermal interface materials, foams, and gels. Measurement of these material forms is extremely difficult and currently not a capability at NIST. Since TOPS provides a direct measurement of thermal conductivity, there is no analysis requirement to input assumed values for heat capacity, density, or sample dimensions, a typical limitation of other thermal conductivity measurement methods. Other thermal conductivity measurement methods also require the sensor to be in direct contact with the sample, leading to parasitic thermal contact errors. TOPS eliminates the need for extensive sample preparation like dimensional shaping or surface polishing, meaning rough surface materials like ceramics, foams, and additively manufactured materials of a variety of shapes and sizes can be readily measured.
The National Institute of Standards and Technology (NIST) intends to negotiate on a sole source basis, under the authority of FAR 13.106-1(b) (1), with Laser thermal located in Charlottesville, VA, for the purchase of this equipment.
Sole Source justification is based on the following:
Based on extensive market research, Laser Thermal Analysis, Inc. is the only vendor with a product that meets our minimum requirements. Other vendors were identified that have products capable of measuring solids, liquids, pastes, gels, and foams, but they do not cumulatively meet our requirements of providing non-destructive, non-contact, spatially resolved, direct measurement of thermal conductivity and for the minimum specified sample dimensions.
Laser thermal has stated “Laser Thermal has no authorized distributors for the Thermo-Optical Plane Source (TOPS) tool. Additionally, TOPS is a proprietary instrument with a patent application pending with the United States Patent and Trademark Office.”
NAICS code for this requirement will be 334516– Analytical Laboratory Instrument Manufacturing with a small business size of 1000 employees.
No solicitation package will be issued. This notice of intent is not a request for competitive quotations; however, interested parties who regard themselves as capable of fulfilling these requirements are invited to submit responses to the Government for consideration. Responses received by 1:00AM EST on 07/07/25 will be considered by the Government.
A determination by the Government not to compete the proposed acquisition based upon responses to this notice is solely within the discretion of the Government. Information received will normally be considered solely for the purpose of determining whether to conduct a competitive procurement.
Inquiry POCs
Erik.Fycklund@nist.gov Primary
Donald.Graham@nist.gov Secondary
National Institutes of Standards and Technologies NIST Acquisition Management Division (AMD)
100 Bureau Dr. Mail Stop 1640
Gaithersburg, MD. 20899-1640
Publications
Every notice SAM.gov issued under this solicitation number, oldest first. Each is a separate record on SAM.
June 20, 2025
Special notice
Due July 7, 2025 at 1:00 PM EDT. SAM.gov, notice 428c8f03a72e49449562accf485e2e86
1 revision on SAM
- June 20, 2025: publish, special notice