# Cardiac Output Monitoring System

Canonical: https://sealedbidder.com/opportunities/ht942626rficoms

- Solicitation number: HT9426-26-RFI-COMS
- Notice type: Special notice
- Status: Closed. Deadline was July 27, 2026 at 5:00 PM EDT
- Department: Department of Defense
- Agency: Defense Health Agency (Dha)
- Contracting office: Defense Health Agency (Dha) (n-DEFENSEHEALTHAGENCYDHA)
- NAICS: 339113 Surgical Appliance and Supplies Manufacturing
- Product or service code: 6515 Medical and Surgical Instruments, Equipment, and Supplies
- Small business size standard: 800 employees (13 CFR 121.201, NAICS 339113)
- Place of performance: JBSA Ft Sam Houston, Texas
- Who can bid, by location: Anywhere in the US. Location does not limit who can bid: a product shipped to the buyer, research at the performer's facility, or work at the contractor's site. A state shown is the delivery point.
- County: Bexar County (FIPS 48029). https://sealedbidder.com/counties/bexar-county-tx-48029
- First posted: July 22, 2026
- Last posted: July 22, 2026
- SAM.gov: https://sam.gov/workspace/contract/opp/2370a4c3176943d8a4d9b1fa35cab395/view

## Reading this notice

- Kind of notice: Special notice. Industry days, draft requirements, program announcements and other notices that are not a solicitation.

## Description

This is a REQUEST FOR INFORMATION (RFI) ANNOUNCEMENT ONLY. In accordance with Revolutionary Federal Acquisition Regulation Overhaul 15.101(c), the Government does not intend to award a contract based on this RFI or otherwise pay for information requested. The information will be treated as information only and not as a proposal. This is not a solicitation announcement nor a request for quotes and does not obligate the Government to award a contract. Requests for a solicitation will not receive a response. Responses to this RFI must be in writing.

The purpose of this RFI announcement is to conduct market research to make appropriate acquisition decisions and to gain knowledge of potential qualified offerors interested and capable of providing the &ldquo;Cardiac Output Monitoring System&rdquo; described below. Documentation of technical expertise must be presented in sufficient detail for the Government to determine that your company possesses the necessary functional area of expertise and experience to fulfill the requirement.

The Defense Health Agency Contracting Activity, 808 Schreider Street, Fort Detrick, MD 21702, is seeking to understand the capabilities of potential qualified Contractors who can provide the supplies and/or services for the following:

Minimum Essential Characteristics: Cardiac Monitoring Platform

The United States Army Institute of Surgical Research (USAISR) seeks a device that can monitor cardiac output (CO) and derive hemodynamic parameters in a hospital setting. The USAISR requires a CO monitoring device that utilizes a Swan-Ganz catheter to monitor hemodynamic parameters. The device&rsquo;s proprietary coded output must be compatible with USAISR&rsquo;s unique internal data logging systems. This is required to maintain historic and current protocol data continuity and workflow. Inability to meet these requirements will inhibit comparisons of current and future studies to USAISR historical data resulting in loss of data, and efforts.

The CO Monitoring Platform shall have the following specifications:

	Ability to measure the following: mixed venous oxygen saturation (SvO2), CO, stroke volume, systemic vascular resistance, right ventricular ejection fraction, right ventricular end diastolic volume, pulmonary artery pressure, and pulmonary artery occlusion pressure.

	Utilization of the Edwards Lifesciences 7.5 FR Swan-Ganz CCOmbo V set or a Swan-Ganz of equal functionality and performance that provides continuous monitoring of SvO2 and measures continuous calculation of cardiac output by using thermodilution principles.

	Equipment dimensions equal to or smaller than 342 x 388 x 208 mm (height x width x depth)

	Equipment shall weigh less than 22 lbs

	Equipment ingress protection of at least IPX1

	Display size of at least 15 inches.

	At least two (2) COM1 serial port connectors, RS-232

	Operational environmental specifications:

		Temperature range: 0 to 37℃

		Relative humidity: 10-90% non-condensing

		Altitude: 0 to 3,000 meters

	Transportation environmental specifications:

		Temperature range: -18 to 45℃

		Relative humidity: 20 to 90% non-condensing

		Altitude: Maximum 6,096 m for up to 8 hours

	Monitor battery must have these technical characteristics:

		Output voltage (nominal) 14.4 V

		Maximum discharge current 4.096 A (8.5 A at 25℃)

		Cells 8 x Li-Ion (lithium ion)

	Monitoring platform technical specifications:

		Input/Output

			Projective capacitive touch screen

			At least two (2) RS-232 serial ports: Maximum data rate = 57.6 kilo baud

			At least three (3) USB ports: 2 x USB 3.0; 1 x USB 2.0

			At least one (1) RJ-45 ethernet port

			At least one (1) HDMI port

		 Heart Rate average display

			Response time: 40 seconds for step increase from 80 to 120 beats per minute (BPM), 29 seconds for step decrease from 80 to 40 BPM.

			CO monitoring on: Averaging time: Time between CO measurements (3 to 21 minutes); Update rate: Approximately 1 minute; Response time: 175 seconds for step increase from 80 to 120 BPM, 176 seconds for step decrease from 80 to 40 BPM.

		Electrical

			Rated supply voltage: 100 to 240 Vac; 50/60 Hz

			Rated input: 1.5 to 2.0 Amps

			Fuses: T 2.5AH, 250V; High breaking capacity; Ceramic

		Alarm

			Sound pressure level: 45 to 85 dB(A)

	Utilization of a Swan-Ganz patient cable with the following parameter measurement specifications:

		Continuous Cardiac Output

			Range: 1 &ndash; 20 Liter/minute

			Reproducibility: &plusmn;6% or 0.1 Liter/minute, whichever is greater

			Maximum thermal filament surface temperature 48℃

		Intermittent (bolus) Cardiac Output

			Range: 1 &ndash; 20 Liter/minute

			Reproducibility: &plusmn;3% or 0.1 Liter/minute, whichever is greater

		Blood temperature

			Range: 15 to 45℃

			Accuracy: &plusmn;3℃

	Roll Stand:

		Compatible rolling stand with bracket and plates for cardiac output monitoring system

			Base Dimensions: 12-15 inches wide x 10-12 inches deep to allow

			Height: Non-specific but need to be adjustable

	Warranty and Product Support:

		12-month standard warranty

		24-month product support and protection plan

The North American Classification System code for this acquisition is 339113 &ndash;

 Surgical Appliance and Supplies Manufacturing (Size Standard: 800 Employees).

Notice to potential offerors: All offerors who provide goods or services to the United States Federal Government must be registered in the System for Award Management (SAM) at www.sam.gov and complete online representations and certifications in SAM.

All interested parties shall submit information by email to both Tammy McRae and Justin Pickett at tammy.n.mcrae.civ@health.mil and justin.h.pickett.civ@health.mil, no later than 5:00 PM Eastern Time on 27 July 2026.

## Open now in the same industry

19 open in NAICS 339113.

- 36C257-27-AP-0953 Prosthetic STX 671 O2 Equipment (VA-27-00016458) (36C25727Q0104), due November 30, 2026 at 2:00 PM EST. https://sealedbidder.com/opportunities/36c25727q0104
- SOURCES SOUGHT NOTICE FOR HEAD, SHOULDER, AND UPPER-EXTREMITY POSITIONING ACCESSORIES FOR THE MAQUET MEERA 7200 OPERATING TABLE (36C26027Q0130), due November 27, 2026 at 3:00 PM EST. https://sealedbidder.com/opportunities/36c26027q0130
- VALVE,BREATHING APP (N0010427QLA17), due November 27, 2026 at 3:30 PM EST. https://sealedbidder.com/opportunities/n0010427qla17

## Publications

- July 22, 2026: Special notice, due July 27, 2026 at 5:00 PM EDT. Notice 2370a4c3176943d8a4d9b1fa35cab395. https://sam.gov/workspace/contract/opp/2370a4c3176943d8a4d9b1fa35cab395/view

## Points of contact

- Tammy McRae, tammy.n.mcrae.civ@health.mil, 3016193771

---
Source: SAM.gov Contract Opportunities bulk extract. Confirm deadlines on SAM.gov before responding. Cite https://sealedbidder.com/opportunities/ht942626rficoms.