Combined synopsis and solicitation
Freeze Dry System- Brand Name or Equal -Brand Name and Equal (BNE)
FDA-RFQ-CBER-2026-135296
Food and Drug Administration, FDA Office of Acq Grant SVCS. Analytical Laboratory Instrument Manufacturing.
Response deadline
September 28, 2026 at 12:00 PM EDT
Closed 59 days ago. Posted September 24, 2026. Scheduled to archive October 13, 2026.
Get the next one like this from Food and Drug Administration by email, the day it is posted.Find my opportunities
This one closed 59 days ago
53 like it are open now in analytical laboratory instrument manufacturing
- Notice of Intent - Illumina MiSeq i100 series instruments
National Institutes of Health, National Institutes of Health - CC
PresolicitationNAICS 334516Ship to Bethesda, MDNOI-OLAO-CC-27-000836Due tomorrowNov 27 - Graphite Furnace Atomic Absorption Spectrometer
National Institute of Standards and Technology, Department of Commerce NIST
Sources soughtNAICS 334516Ship to Gaithersburg, MDNIST-MML-FY27-SS01Due in 2 daysNov 28 - Intent to Solicit Only One Source for Six (6) Twinleaf Pulsed Pump Magnetometer (part number: PPM-RS422) and one (1) Twinleaf 8 Channel Synchronization Hub (part number: HUB-RS422-8) manufactured by Twinleaf.
Department of the Army, W6QK ACC-APG Adelphi
SolicitationNAICS 334516Ship to MDW911QX27QA015Due in 5 daysDec 1
All 53 open in analytical laboratory instrument manufacturing
Reading this notice
Worked out from the publications and the award on this page.
- Kind of notice
- Combined synopsis and solicitation. A streamlined commercial-item buy where the notice itself is the solicitation. Glossary
- Who may compete
- SAM.gov lists no set-aside on this notice, so it is not reserved for small businesses. Small, under NAICS 334516 (analytical laboratory instrument manufacturing), means no more than 1,000 employees, counted with affiliates. What a set-aside is
- Time to respond
- Bidders had 4 days from the day the solicitation was posted, September 24, 2026, to the deadline.
Size standard from 13 CFR 121.201. The solicitation itself is what binds: check the code and standard it states.
Description
As published on SAM.gov.
Notice of Combined Synopsis/Solicitation. This is a combined synopsis/solicitation for commercial products or commercial services prepared in accordance with part 12. This announcement constitutes the only solicitation; quotes are being requested, and a separate written solicitation will not be issued. The NAICS for this requirement is: 334516. The government is not obligated to make an award. The Government is not obligated to pay costs incurred by the Contractor during quote preparation and quote submission.
Background: The U.S. Food and Drug Administration (FDA) laboratory requires a benchtop freeze dryer (lyophilizer) system to support ongoing scientific research and sample preservation activities. Lyophilization (freeze-drying) is a critical process used to preserve the stability and integrity of biological samples, pharmaceuticals, and other moisture-sensitive materials by removing water content through sublimation under vacuum. The required system must achieve a collector temperature of -84 °C or colder to accommodate samples with lower freezing points, including those containing organic solvents. The system must be configured for 230V/50-60 Hz operation consistent with the FDA laboratory's electrical infrastructure.
Minimum Performance Requirements/Salient Characteristics:
This is a Brand Name or Equal requirement. Products other than the aforementioned are acceptable but shall meet all the salient characteristics contained in the Minimum Technical Requirements below:
All of the following system requirements are minimum requirements. Equivalent requirements that differ from these minimum requirements shall be justified by the proposing vendor and evaluated prior to purchase:
General
The contractor shall provide One (1) benchtop freeze dryer (lyophilizer) system meeting the following minimum technical specifications, including all required accessories, tubing, clamps, adapters, and standard warranty necessary for full system assembly and operation:
• Collector Temperature: -84 °C or colder (to accommodate samples containing organic solvents such as acetonitrile or methanol)
• Collector Capacity: Approximately 2.5 liters of ice capacity before defrost cycle
• Condenser Type: Stainless steel or PTFE-coated coils (corrosion-resistant construction)
• Drying Chamber: 12-port drying chamber or manifold for simultaneous multi-sample processing
• Accessories: Initial set of flasks (150 mL and 300 mL), all necessary tubing, clamps, and adapters required for full system assembly
• Electrical: 230V, 50/60 Hz configuration
Place of Performance:
The prospective vendor should provide insured shipping of equipment to:
U.S. Food and Drug Administration,
Attn: Kyung Sung
10903 New Hampshire Ave, Building 52/72. Room 3248
Silver Spring, MD 20993
Period of Performance:
Performance begins on the date of formal government acceptance and continues through the end of the manufacturer's warranty period.
Quotes are due by email on or before September 28 2026, by 12:00 P.M. (Eastern Standard Time) and should be sent to brandon.rafus@fda.hhs.gov.
Attachments
Files listed on SAM.gov. The ones we hold a copy of download from here; the rest open SAM's download, which needs no account.
- SOW- CBER-2026-135296- Final.pdf112 KB, September 24, 2026
Publications
Every notice SAM.gov issued under this solicitation number, oldest first. Each is a separate record on SAM.
September 24, 2026
Combined synopsis and solicitation
Due September 28, 2026 at 12:00 PM EDT. SAM.gov, notice 47c131633cca4e6ca6beda277e89d493
1 revision on SAM
- September 24, 2026: publish, combined synopsis and solicitation
Points of contact
- Brandon Rafusbrandon.rafus@fda.hhs.gov