Presolicitation
Automatic Liquid Sampler for Agilent 8890 Gas Chromatograph
ARS_1172105
Food Safety and Inspection Service, Usda, Fsis, Oas PCMB. Analytical Laboratory Instrument Manufacturing.
Awarded
$24,365.21 obligated so far on USAspending
Get the next one like this from Food Safety and Inspection Service by email, the day it is posted.Find my opportunities
This one was awarded in 2026
61 like it are open now in analytical laboratory instrument manufacturing
- Calibrate PAMAS SBSS
Department of the Navy, Portsmouth Naval Shipyard GF
SolicitationNAICS 334516San Diego, CAN3904027Q5950Due todayNov 25 - ChemiDoc Go Imaging System
National Institutes of Health, National Institutes of Health Olao
Combined synopsis and solicitationNAICS 334516Ship to Hercules, CA75N98026Q00083Due in 3 daysNov 28 - Spectrum Analyzer
Department of the Navy, Naval Surface Warfare Center
Combined synopsis and solicitationNAICS 334516Ship to Norco, CAN6426727Q4009Due in 7 daysDec 2
All 61 open in analytical laboratory instrument manufacturing
Reading this notice
Worked out from the publications and the award on this page.
- Kind of notice
- Presolicitation. Advance notice that a solicitation is coming, usually within a few weeks. Glossary
- Who may compete
- SAM.gov lists no set-aside on this notice, so it is not reserved for small businesses. Small, under NAICS 334516 (analytical laboratory instrument manufacturing), means no more than 1,000 employees, counted with affiliates. What a set-aside is
- How it ended
- Awarded to Technical Communities, Inc. on September 18, 2026, 8 days after offers were due. The contracting office reported 2 offers (competed under simplified acquisition procedures).
Size standard from 13 CFR 121.201. The solicitation itself is what binds: check the code and standard it states.
Description
As published on SAM.gov.
The U.S. Department of Agriculture (USDA), Agricultural Research Service (ARS), Roman L. Hruska U.S. Meat Animal Research Center (USMARC) in Clay Center, NE, intends to issue a solicitation for a firm-fixed-price contract to procure brand-name Agilent accessories, specifically an Automatic Liquid Sampler, Split/Splitless Inlet, and a Flame Ionization Detector (FID) compatible with an existing Agilent 8890 Gas Chromatograph.
This procurement is conducted under the authority of FAR 13.106-1(b) for brand-name specific items. The required products must seamlessly integrate and operate with the government’s existing Agilent 8890 infrastructure to support ongoing environmental sample analysis.
Salient Characteristics & Technical Requirements:
ITEM No. 1: Split/Splitless Inlet (Agilent Part No. G3548A) – Quantity: 1
• Must be fully compatible with the Agilent 8890 gas chromatograph.
• Suitable for all capillary columns (50 to 530 μm id).
• Capable of split ratios up to 12,500:1 to prevent column overload.
• Features splitless mode for trace analysis with easily accessible pressure-pulsed splitless performance.
• Maximum operating temperature of at least 400 °C.
• EPC available in two pressure ranges: 0 to 100 psig for columns ≥0.200 mm diameter, and 0 to 150 psig for columns <0.200 mm diameter.
• Gas saver mode to reduce consumption without compromising performance.
• Electronic septum purge flow control to eliminate ghost peaks.
• Total flow setting ranges: 0 to 500 mL/min N2; 0 to 1,250 mL/min H2 or He; 0 to 200 mL/min argon/methane.
• Built-in standard turn-top inlet sealing system for rapid, tool-less liner changes.
• Chemically deactivated weldment and weldment insert.
ITEM No. 2: Autosampler / Automatic Liquid Sampler (Agilent Part No. G4567A) – Quantity: 1
• Must be fully compatible with the Agilent 8890 gas chromatograph.
• Supports standard 2 mL vials and micro vial inserts.
• Minimum capacity of 50 samples.
• Sample discrimination ≤ 10%.
• Fully programmable dispense rate, draw rate, and injection rate.
• Fast injection speeds executed in less than 100 ms.
• Support for 250 and 500-μL syringes via an optional Large Volume Syringe Carriage.
• Equipped with sensors to detect the presence of the Large Volume Syringe Carriage.
• Features a user-changeable syringe carriage and self-aligning injector.
ITEM No. 3: Flame Ionization Detector (FID) (Agilent Part No. G3532A) – Quantity: 1
• Must be fully compatible with the Agilent 8890 gas chromatograph.
• Responds broadly to organic compounds.
• Minimum detectable level (for tridecane) <1.2 pg C/s.
• Linear dynamic range >10^7 (±10%) utilizing a full-range digital data path.
• Data rates up to 1,000 Hz to handle peaks as narrow as 5 msec at half height.
• Standard electronic pneumatic control for three gases: Air (0-800 mL/min), H2 (0-100 mL/min), and Make-up gas (N2 or He, 0-100 mL/min).
• Flameout detection and automatic reignition capabilities.
• Maximum operating temperature of at least 450 °C.
• Configured to be mounted as a third detector on the left side of the GC system.
Installation and Performance Requirements:
The contractor is responsible for the delivery, complete physical installation, and operational verification testing of all three items. The contractor must coordinate schedules directly with the government Point of Contact (POC) to establish a mutually agreeable installation timeline. All equipment must be fully operational and verified by the contractor within the timeline specified below.
Delivery and Place of Performance:
All items must be delivered, installed, and deemed operational within ninety (90) calendar days after contract award.
Delivery Location:
USDA ARS USMARC
844 Road 313
Clay Center, NE 68933
This notice does not constitute a Request for Quotation (RFQ) or Request for Proposal (RFP). No response is required for this pre-solicitation notice unless requesting additional capability verification.
The contract, on USAspending
Federal procurement data the awarding office reported to FPDS, matched to this solicitation by its number.
- Recipient
- Technical Communities, Inc.
- UEI
- MCQUYM6GHNJ5
- CAGE
- 1RPN6
- Vendor location
- San Mateo, CA
- Contract
- 123A9426P0049, purchase order
- Obligated
- $24,365.21
- Actions
- 1 between September 18, 2026 and September 18, 2026
- Competition
- Competed Under SAP, 2 offers received
- Set-aside reported
- No Set Aside Used.
- Described as
- Automatic Liquid Sampler for Agilent 8890 Gas Chromatograph and Accompanying Additions
- Match
- solicitation number ARS1172105 equals the FPDS solicitation identifier; same awarding office 123A94 (high confidence)
- Record
- USAspending award page
Publications
Every notice SAM.gov issued under this solicitation number, oldest first. Each is a separate record on SAM.
September 8, 2026
Presolicitation
Due September 10, 2026 at 8:00 AM EDT. SAM.gov, notice 2a41976ac1034a819693d19db7ed689d
2 revisions on SAM
- September 8, 2026: publish, presolicitation
- September 8, 2026: publish, presolicitation
September 8, 2026
Presolicitation
Due September 10, 2026 at 8:00 AM EDT. SAM.gov, notice 79209038355149e0be0d449f525c7269
2 revisions on SAM
- September 8, 2026: publish, presolicitation
- September 8, 2026: publish, presolicitation
Points of contact
- Angela Careyangela.carey@usda.gov3018377781