# Acquisition of an Applied Biosystems QuantStudio 3 Real-Time PCR System

Canonical: https://sealedbidder.com/opportunities/75n98026q00714

- Solicitation number: 75N98026Q00714
- Notice type: Combined synopsis and solicitation
- Status: Closed. Deadline was July 20, 2026 at 9:00 AM EDT
- Department: Department of Health and Human Services
- Agency: National Institutes of Health
- Contracting office: National Institutes of Health Olao (75N980)
- NAICS: 334516 Analytical Laboratory Instrument Manufacturing
- Product or service code: 6640 Laboratory Equipment and Supplies
- Small business size standard: 1,000 employees (13 CFR 121.201, NAICS 334516)
- Place of performance: Bethesda, Maryland
- Who can bid, by location: Anywhere in the US. Location does not limit who can bid: a product shipped to the buyer, research at the performer's facility, or work at the contractor's site. A state shown is the delivery point.
- County: Montgomery County (FIPS 24031). https://sealedbidder.com/counties/montgomery-county-md-24031
- City: Bethesda. https://sealedbidder.com/cities/bethesda-md-2407125
- First posted: July 9, 2026
- Last posted: July 9, 2026
- SAM.gov: https://sam.gov/workspace/contract/opp/3200a68293ae4a1e8cf459207cf69cc5/view

## Reading this notice

- Kind of notice: Combined synopsis and solicitation. A streamlined commercial-item buy where the notice itself is the solicitation.
- Who may compete: SAM.gov lists no set-aside on this notice, so it is not reserved for small businesses. Small, under NAICS 334516 (analytical laboratory instrument manufacturing), means no more than 1,000 employees, counted with affiliates.
- Time to respond: Bidders had 11 days from the day the solicitation was posted, July 9, 2026, to the deadline.

## Description

Requirements:

To ensure continuity of research, NIAMS proposes the acquisition of an Applied Biosystems QuantStudio 3 Real-Time PCR System. qPCR is a core technology used for nucleic acid amplification, quantification, and analysis and supports a wide range of research applications, including gene expression studies, pathogen detection, transcript quantification, and validation of experimental results. The QuantStudio 3 is the closest replacement to the current StepOnePlus systems, allowing researchers to maintain established workflows and consistency in experimental design and data interpretation. The QuantStudio 3 offers several operational advantages, including user-friendly software, compatibility with both Mac and PC platforms, touchscreen operation, and the ability to function as a standalone instrument or in conjunction with a computer. Its thermal block design is similar to the existing systems, minimizing the learning curve for users and supporting a broad range of genomic applications. In addition, the manufacturer offers trade-in credit for the current instruments, reducing the overall acquisition cost while providing access to a fully supported platform with available parts and technical support. Acquiring the QuantStudio 3 at this time will help prevent interruptions to research activities, maintain reliable and consistent data generation, and ensure that NIAMS laboratories have access to a supported and dependable qPCR platform.

4.1 PRODUCT FEATURES/SALIENT CHARACTERISTICS The contractor shall provide one Applied Biosystems QuantStudio 3 Real-Time PCR System or an equal system that meets all the following minimum requirements. 4.1.1. System Features/Benefits The system shall: &bull; Utilize a platform providing 1 TB or storage space for data sharing at user discretion &bull; be used with online scientific apps &bull; come with desktop-based software, allowing single-plate run setup and analysis.

4.1.2. Technical specifications &ndash; Instrument The instrument shall: &bull; have a 96 well 0.1 ml plate block format and the option to run fast protocols <30 minutes &bull; not require a field service engineer for set up by end user. Set up should be easy for end user &ndash; directly out of the box. &bull; not require a computer for installation &bull; use an RNase P verification plate to verify performance &bull; not require dye calibration upon installation &bull; have dimensions fo 27 x 50 x 40 cm and be designed to maximize horizonal and vertical bench space &bull; be capable of analyzing 96 samples simultaneously in various formats incluing both 96-well plates, 8-tube strips and individual tubes &bull; the 96-well block should be made of at least 3 separated Peltier elements &ndash; VeriFlex Blocks. The maximum temperature difference programmable across the block should be 10 degrees C. 5 degrees C should be the maximum difference in temperature allowed between adjacent blocks. &bull; Support reaction volumes of 10-30 uL on 0.1mL the 96-well block &bull; Be factory calibrated with FAM, SYBR Green 1, VIC, NED, ABY, JUN, TAMRA, and ROX dyes &bull; Have melt or dissociation curves that take place in steps with deltaT varying between 0.015 degrees C and 3.66 degrees C &bull; Have a maximum block ramp rate of 9.0 degrees C/sec for the 0.1 mL block &bull; Have a block with a temperature range of 4-99.9 degrees celcius &bull; Have a heated cover with temperature range of 50-110 degreesC 2

&bull; Have a long-lasting bright white LED source with median lifetime of at least 60,000 hours &bull; Have a CMOS imager for data acquisition and collects data for each filter combination in less than 2 seconds &bull; be used with 120-240 V outlets without modification or special plug requirements &bull; Include a heated lid assembly to heat tops of sample plates and minimize reaction evaporation.

 4.1.3 Technical specifications &ndash; Instrument onboard computer The system shall: &bull; Possess the capability to run in stand-alone mode, without an attached computer &bull; Include a touchscreen interface and store protocols for quick starts with no computer needed &bull; Have a touchscreen interface that enables viewing amplification plots and filtering them by sample, target and task &bull; Have a touchscreen interface that facilitates uploading of protocols or downloading of run files via a USB, local network and the Connect platform &bull; Have a touchscreen interface that allows uploading of protocols and run files directly to the Connect platform &bull; Contain onboard storage with memory capacity of 10 GB for storing 1,000-5,000 standard absolute quantification run files, and be designed for run protection in case of power outages &bull; Be able to be customized through the touchscreen for time zones and service/calibration notification settings &bull; Have a touchscreen interface that supports the creation of local user accounts with PIN-protected access that can be linked to the Connect platform to online accounts on thermofisher.com &bull; Have a touchscreen interface that allows creation and running of protocols as guests &bull; Have a touchscreen interface that allows for editing of protocols at all stages of the reaction cycle as well as block settings, temperature settings duration, ramp rate, etc. &bull; Have a touchscreen interface that allows a template to be loaded from a USB, local network or the Connect platform &bull; Have a touchscreen interface that allows user to enter reagent information &bull; Have a touchscreen interface that allows a run to be paused and during that pause for the block temperature to be set as well as for the drawer to be opened and closed &bull; Have a touchscreen interface that allows editing of sample and target assignments &bull; Have a touchscreen interface that allows the repeat of a previous run 3 &bull; Have a touchscreen interface that allows users to perform dye calibrations and view results on the touchscreen &bull; Have a touchscreen interface that allows RNase P verification runs and automatically provides pass/fail results &bull; Have a touchscreen interface that allows a user to run a self-verification test &bull; Have a touchscreen interface that allows the user to view the cure QC assessment after calibration and RNase P plate runs are finished &bull; Have a touchscreen interface that allows progress monitoring.

 4.1.4 Performance Requirements The instrument shall: &bull; Complete 40-cycle real-time PCR runs using fluorogenic 5&rsquo; nuclease assay and Fast chemistry in a standard 96-well plate in less than 30 minutes as well as being able to run in standard ramping mode with standard chemistry &bull; Include an OptiFlex system that combines four coupled excitation, and emission filter sets to allow multiplexing up to four unique wavelengths &bull; Support at least two reaction chemistries-TaqMan&reg; Probes and SYBR Green dye &bull; Have real-time quantitative PCR installation specifications to be able to distinguish between 5,000 and 10,000 template copies at a 99.7% confidence level.

 4.1.5 Software Requirements The software shall: &bull; Allow flexible customization of run protocols &bull; Allow the creation of secure templates that are locked with a password to limit modification of run setup &bull; Provide run monitoring of connected instruments &bull; Include preoptimized run protocols for applications like standard curve, relative standard curve, relative quantification of gene expression, and SNP genotyping with TaqMan&reg; or SYBR Green chemistries &bull; Include a copy of free software to analyze data for absolute and relative quantification, SNP detection and other assays. &bull; Collect information on both TaqMan and SYBR assays in the same run &bull; Have a genotyping optimization tool that uses real-time data &bull; Provide raw fluorescence data and multicomponent data for troubleshooting &bull; Allow conversion of protocols for the Applied Biosystems StepOnePlus and 7500 to the QuantStudio 3 system protocols &bull; Allow batch analysis of up to 500 experiments simultaneously &bull; Can control the instrument and analyze data from a remote computer on the same network 4 &bull; Provide status monitoring of connected instruments and real-time plots STATEMENT OF NEED (SON) &bull; Have the ability to send user notifications when a run is started, stopped or has an error &bull; Allow registered users of the Connect platform to monitor any connected instrument &bull; Allow users to run an assay plate without performing plate setup prior to beginning of run and to correct plate set up after the run &bull; Allow export of all run results to excel and save graphs as .jpg files &bull; Provide the ability to import a standard curve from one plate for analysis of another plate of the same instrument type &bull; Allow for the selection of endogenous controls for gene expression experiments &bull; Include statistical analysis tools including box-whisker plots, scatter plots and heat maps &bull; Allow preselection of data to export and the location of the exported file &bull; Allow for command-line functions for batch exporting of analyzed run files &bull; Allow users to save predefined analysis settings for auto-exporting run data into format of choice &bull; Contain online high-resolution melt application so user can specify number of clusters/variants and save defined controls and analysis setting.

 4.1.6 Instrument Ecosystem The instrument ecosystem shall: &bull; Be part of a desktop ecosystem that consists of a suite of qPCR applications. Instrument data can be grouped (multi-plate) for relative quantitation, SNP genotyping, standard curve, HRM analysis, and presence/absence analysis. &bull; Be part of the Connect platform &bull; Be able to connect to the Connect platform for run setup download and data uploads &bull; Allow a run to be monitored by a web browser or mobile device &bull; Allow batch-analysis of up to 500 experiments simultaneously through an online relative quantitation module &bull; Provide an endogenous control selection tool for gene expression experiments in both the desktop and online quantitation software &bull; Contain statistical anaylsis tools with both the desktop and online quantitation software &bull; Contain a genotyping optimization tool that uses real-time data to pick the best cycle number for calling genotype in both the desktop and online genotyping software &bull; Allow the user to specify number of clusters/variants and save defined controls and analysis settings with both desktop and online software.

 4.1.7.  Design 5 The system shall: &bull; Have an application programming interface (API) included that allows programming routines to be developed for controlling the instrument &bull; Be cUL, UL, CE, C-TICK, and WEEE compliant &bull; Be developed and manufactured in a facility with quality system requirements that comply with ISO 13485 standards and ISO 9001 standards 4.1.8 Gene expression profiling &bull; Users shall be able to select from more than 61,000 predesigned human, mouse and rat Applied BiosystemsTMTaqMan&reg; Assays.

 4.1.9 Able to accept the trade-in of Model STEPONEPLUS. Serial number 272008444 4.1.10 Must be compatible with protocols from Applied Biosystems StepOnePlus Real Time PCR (qPCR) allowing for consistency in experimental set-up 4.2 DELIVERY / INSTALLATION The contractor shall deliver and install the equipment within thirty (30) calendar days after receipt of award.

## Attachments

- [Requirements rev8.docx](https://sam.gov/api/prod/opps/v3/opportunities/resources/files/5e8b61c585064d978a2e3b5234727ebd/download)

## Open now in the same industry

53 open in NAICS 334516.

- Notice of Intent - Illumina MiSeq i100 series instruments (NOI-OLAO-CC-27-000836), due November 27, 2026 at 12:00 PM EST. https://sealedbidder.com/opportunities/noiolaocc27000836
- Graphite Furnace Atomic Absorption Spectrometer (NIST-MML-FY27-SS01), due November 28, 2026 at 2:00 PM EST. https://sealedbidder.com/opportunities/nistmmlfy27ss01
- Intent to Solicit Only One Source for Six (6) Twinleaf Pulsed Pump Magnetometer (part number: PPM-RS422) and one (1) Twinleaf 8 Channel Synchronization Hub (part number: HUB-RS422-8) manufactured by Twinleaf. (W911QX27QA015), due December 1, 2026 at 11:59 AM EST. https://sealedbidder.com/opportunities/w911qx27qa015

## Publications

- July 9, 2026: Combined synopsis and solicitation, due July 20, 2026 at 9:00 AM EDT. Notice 3200a68293ae4a1e8cf459207cf69cc5. https://sam.gov/workspace/contract/opp/3200a68293ae4a1e8cf459207cf69cc5/view

## Points of contact

- John Burton, burtonjo@mail.nih.gov

---
Source: SAM.gov Contract Opportunities bulk extract. Confirm deadlines on SAM.gov before responding. Cite https://sealedbidder.com/opportunities/75n98026q00714.