# Multisizer 4e Coulter Counter

Canonical: https://sealedbidder.com/opportunities/75n98026q00434

- Solicitation number: 75N98026Q00434
- Notice type: Presolicitation
- Status: Closed. Deadline was June 3, 2026 at 9:00 AM EDT
- Department: Department of Health and Human Services
- Agency: National Institutes of Health
- Contracting office: National Institutes of Health Olao (75N980)
- NAICS: 334516 Analytical Laboratory Instrument Manufacturing
- Product or service code: 6640 Laboratory Equipment and Supplies
- Small business size standard: 1,000 employees (13 CFR 121.201, NAICS 334516)
- Place of performance: Maryland
- Who can bid, by location: Anywhere in the US. Location does not limit who can bid: a product shipped to the buyer, research at the performer's facility, or work at the contractor's site. A state shown is the delivery point.
- County: Montgomery County (FIPS 24031). https://sealedbidder.com/counties/montgomery-county-md-24031
- First posted: May 22, 2026
- Last posted: May 22, 2026
- SAM.gov: https://sam.gov/workspace/contract/opp/1a8af077f9a74b35963e4c47cfe527cf/view

## Reading this notice

- Kind of notice: Presolicitation. Advance notice that a solicitation is coming, usually within a few weeks.
- Who may compete: SAM.gov lists no set-aside on this notice, so it is not reserved for small businesses. Small, under NAICS 334516 (analytical laboratory instrument manufacturing), means no more than 1,000 employees, counted with affiliates.
- What comes next: No solicitation was published under this number after the presolicitation of May 22, 2026. An agency sometimes issues the solicitation under a new number.

## Description

The required Multisizer provides improved functionality and analysis capabilities, in addition to allowing continued measurement of cell size, counts, and distribution using the Coulter method across multiple clinical research study protocols.

                                                  Specifications & Requirements

                        The product shall include/use:

	Coulter method of measurement using patented digital pulse processing technology

	Automatic aperture blockage detection and clearing during runs and sample analysis

	Code scanner to scan bar codes on samples, electrolytes, aperture tubes, and sample beakers

	Fluid level monitoring for solutions and waste

	Selectable stirrer speed and direction

	Particle sizing range: 0.2 &ndash; 1,600uM

	Measurement range: 2 &ndash; 80% of aperture size

	Measurement linearity: Diameter &plusmn; 1%, Volume &plusmn; 3%

	Dynamic range (accuracy): 1:40 (diameter), 1:64,000 (volume)

	Accurate determination of sample cell number, size, distribution, and concentration

	Particle count, sizing, and volume distribution data, including diameter, volume, and area of number, number percent, number per mL, volume, volume percent, volume per mL, surface area, surface area percent, surface area per mL

	Pulse distribution data, including time, sequence and width for pulse height diameter, pulse height volume, pulse height volt, pulse width, pulse area, average pulse height diameter, average pulse height volume and average pulse width. Number distribution by width.

	Up to 400 size classes or channels with the ability to change number of size classes or channels and width and reprocess as needed without having to re-run the samples

	Time-stamped pulses up to 525,000 per single analysis

	Modes for Total Count (up to 500,000 counts), Modal Count (up to 100,000 counts), and Volumetric Count (up to 2000uL)

	Digital Pulse Processor with proprietary high-speed signal digitalization

	Real time detection of volume changes

	Stirrer with ability to adjust stir speed and direction for even particle distribution in sample suspensions

	100uM and 400uM size aperture tubes, with instrument ability to use apertures ranging from 10uM up to 2000uM size as needed, with aperture tube calibration and verification

	100/200mL beakers, with ability to use additional size beakers and accuvettes, as required by the sample type, sample volume, or other need

	Isoton II Diluent and Coulter Clenz solutions for instrument operation and cleaning, and spigot assemblies for both solutions

	Mixed Standards Kit (L2, L5, L10, L20, L40)

	300uM Particle Standard

	Installation

## Attachments

- [Specifications Requirements rev2.docx](https://sam.gov/api/prod/opps/v3/opportunities/resources/files/9e7db78c24fa4fb18300dfc8a7c2014f/download)

## Open now in the same industry

53 open in NAICS 334516.

- Notice of Intent - Illumina MiSeq i100 series instruments (NOI-OLAO-CC-27-000836), due November 27, 2026 at 12:00 PM EST. https://sealedbidder.com/opportunities/noiolaocc27000836
- Graphite Furnace Atomic Absorption Spectrometer (NIST-MML-FY27-SS01), due November 28, 2026 at 2:00 PM EST. https://sealedbidder.com/opportunities/nistmmlfy27ss01
- Intent to Solicit Only One Source for Six (6) Twinleaf Pulsed Pump Magnetometer (part number: PPM-RS422) and one (1) Twinleaf 8 Channel Synchronization Hub (part number: HUB-RS422-8) manufactured by Twinleaf. (W911QX27QA015), due December 1, 2026 at 11:59 AM EST. https://sealedbidder.com/opportunities/w911qx27qa015

## Publications

- May 22, 2026: Presolicitation, due June 3, 2026 at 9:00 AM EDT. Notice 1a8af077f9a74b35963e4c47cfe527cf. https://sam.gov/workspace/contract/opp/1a8af077f9a74b35963e4c47cfe527cf/view

## Points of contact

- John Burton, burtonjo@mail.nih.gov

---
Source: SAM.gov Contract Opportunities bulk extract. Confirm deadlines on SAM.gov before responding. Cite https://sealedbidder.com/opportunities/75n98026q00434.