Presolicitation, total small business set-aside
Insect rearing chamber
75F40126Q135135
Food and Drug Administration, FDA Office of Acq Grant SVCS. Analytical Laboratory Instrument Manufacturing.
Response deadline
July 7, 2026 at 2:00 PM EDT
Closed 142 days ago. Posted June 30, 2026. Scheduled to archive July 22, 2026.
Get the next one like this from Food and Drug Administration by email, the day it is posted.Find my opportunities
This one closed in 2026
55 like it are open now in analytical laboratory instrument manufacturing
- Notice of Intent - Illumina MiSeq i100 series instruments
National Institutes of Health, National Institutes of Health - CC
PresolicitationNAICS 334516Ship to Bethesda, MDNOI-OLAO-CC-27-000836Due tomorrowNov 27 - Graphite Furnace Atomic Absorption Spectrometer
National Institute of Standards and Technology, Department of Commerce NIST
Sources soughtNAICS 334516Ship to Gaithersburg, MDNIST-MML-FY27-SS01Due in 2 daysNov 28 - Intent to Solicit Only One Source for Six (6) Twinleaf Pulsed Pump Magnetometer (part number: PPM-RS422) and one (1) Twinleaf 8 Channel Synchronization Hub (part number: HUB-RS422-8) manufactured by Twinleaf.
Department of the Army, W6QK ACC-APG Adelphi
SolicitationNAICS 334516Ship to MDW911QX27QA015Due in 5 daysDec 1
All 55 open in analytical laboratory instrument manufacturing
Reading this notice
Worked out from the publications and the award on this page.
- Kind of notice
- Presolicitation. Advance notice that a solicitation is coming, usually within a few weeks. Glossary
- Who may compete
- Total small business set-aside. Only small businesses may compete. Small, under NAICS 334516 (analytical laboratory instrument manufacturing), means no more than 1,000 employees, counted with affiliates. Open Small business opportunities
- What comes next
- No solicitation has been published under this number yet. The presolicitation of June 30, 2026 is the earliest notice of the purchase.
Size standard from 13 CFR 121.201. The solicitation itself is what binds: check the code and standard it states.
Description
As published on SAM.gov.
The associated North American Industry Classification System (NAICS) code for this acquisition is 334516 – Analytical Laboratory Instrument Manufacturing; the applicable Small Business Size Standard is 1000 employees. This requirement is a small business set aside.
Background:
The U.S. Food and Drug Administration (FDA), Center for Biologics Evaluation and Research (CBER), Laboratory of Vector-Borne Viral Diseases (LVVD) requires the purchase of an Insect Rearing Chamber.
The LVVD of Division of Viral Products conducts research on important vector-borne viral pathogens, including dengue, Zika, and chikungunya viruses. We frequently review product submissions that involve live-attenuated vector-borne viruses intended as vaccines. An important consideration with any such product is whether it is capable of being transmitted to insect vectors by the vaccinated patient. In the past, we have been relying on individual reviewer expertise in medical entomology garnered through past experience. With the need for more personnel trained in vector biology and primary research into vector-related products, we have concluded that solely relying on past experience can no longer keep up with the pace of regulatory review and the industry. In order to revitalize the FDA’s expertise in vector-borne disease research, which is incomplete without studying insect vectors, we require a dedicated insect cell incubator for rearing mosquito colonies. Maintaining stable temperature, light cycles, and humidity is critical to the integrity of ongoing research involving insect colonies. To support this work, the laboratory requires the purchase of a dedicated insect rearing chamber capable of maintaining consistent temperature and relative humidity conditions suitable for insect rearing applications.
The Contractor shall provide one (1) insect rearing chamber that meets the following minimum requirements:
- Temperature Control
- Capable of maintaining a temperature setpoint of 27°C.
- Temperature control accuracy of ±0.2°C or better.
- Humidity Control
- Active humidification capable of maintaining 75–80% relative humidity (RH).
- Humidity control accuracy of ±0.3% RH or better.
- Humidity range from ambient conditions to at least 90% RH.
- Lighting: Interior LED lighting with a programmable 24-hour timer capable of simulating day/night (diurnal) cycles appropriate for insect rearing.
- Capacity: Minimum interior capacity of 20 cubic feet.
- Cooling System: Equipped with a cooling system that ensures temperature stability during normal operation and in the event of a component failure.
Installation, Training and Additional System Requirements:
- The components and equipment shall be newly manufactured, not used or refurbished, or previously used for demonstration.
- Offered systems shall be a turn-key solution i.e., the contractor shall be responsible for providing all hardware, components, instruments, computers, software, and that otherwise required to meet these specifications and the FDA’s stated need.
- The systems shall be delivered with all necessary supplies and accessories required for installation and start-up.
- The Contractor shall provide NIST-traceable calibration documentation for the chamber's temperature and humidity sensors. Documentation shall include a minimum three-point calibration certificate and be provided with the equipment upon delivery.
- The system and associated accessories shall include operations and maintenance manuals covering proper operation, routine maintenance, and troubleshooting for the system and controlling software. All manuals and documentation shall be provided in hard copy and/or electronic format.
- The Contractor shall provide all labor, travel, and tools to install the equipment at the address provided below, to include shipping. The contractor shall demonstrate upon installation that the item meets or exceeds all performance specifications. Upon acceptance of the system, the contractor shall provide on-site operator training/familiarization. Such familiarization shall include system operations, calibration, optimization, troubleshooting and basic operational maintenance procedures.
- Systems shall be warranted for not less than one (1) year from FDA acceptance of the system(s). Warranty service shall include troubleshooting capabilities based on complete knowledge of the entire system, immediate access to replacement parts. Phone and email technical support shall be included for a minimum of one (1) year.
Place of Delivery:
FOB Point Destination. All items shall include shipping, handling, installation, and inside delivery to the destination identified herein.
FDA/CBER/OVRR
Building 52-72 room 5353
10903 New Hampshire Avenue
Silver Spring, MD 20993
Period of Performance:
Delivery, installation, calibration and training shall occur within 45 calendar days from date of award.
All responsible sources may submit a quotation, bid, or proposal, as appropriate, which will be considered by the agency
For information regarding this notice, please contact Warren Dutter at 301-796-2486 or email warren.dutter@fda.hhs.gov.
Publications
Every notice SAM.gov issued under this solicitation number, oldest first. Each is a separate record on SAM.
June 30, 2026
Presolicitation
Due July 7, 2026 at 2:00 PM EDT. SAM.gov, notice f71f1ca345e04415839c23ab05436b83
1 revision on SAM
- June 30, 2026: publish, presolicitation
Points of contact
- Warren Dutterwarren.dutter@fda.hhs.gov3017962486