# 3D Imaging Microscope System

Canonical: https://sealedbidder.com/opportunities/75f40126q00383

- Solicitation number: 75F40126Q00383
- Notice type: Combined synopsis and solicitation
- Status: Awarded to Bruker Nano, Inc.
- Department: Department of Health and Human Services
- Agency: Food and Drug Administration
- Contracting office: FDA Office of Acq Grant SVCS (75F401)
- NAICS: 334519 Other Measuring and Controlling Device Manufacturing
- Product or service code: 6640 Laboratory Equipment and Supplies
- Small business size standard: 600 employees (13 CFR 121.201, NAICS 334519)
- Place of performance: Cincinnati, Ohio
- Who can bid, by location: Anywhere in the US. Location does not limit who can bid: a product shipped to the buyer, research at the performer's facility, or work at the contractor's site. A state shown is the delivery point.
- County: Hamilton County (FIPS 39061). https://sealedbidder.com/counties/hamilton-county-oh-39061
- City: Cincinnati. https://sealedbidder.com/cities/cincinnati-oh-3915000
- First posted: September 8, 2026
- Last posted: September 8, 2026
- SAM.gov: https://sam.gov/workspace/contract/opp/44430997898d4b699da81c4f883899c2/view

## Reading this notice

- Kind of notice: Combined synopsis and solicitation. A streamlined commercial-item buy where the notice itself is the solicitation.
- Who may compete: SAM.gov lists no set-aside on this notice, so it is not reserved for small businesses. Small, under NAICS 334519 (other measuring and controlling device manufacturing), means no more than 600 employees, counted with affiliates.
- Time to respond: Bidders had 6 days from the day the solicitation was posted, September 8, 2026, to the deadline.
- How it ended: Awarded to Bruker Nano, Inc. on September 30, 2026, 16 days after offers were due. The contracting office reported 2 offers (competed under simplified acquisition procedures).

## Description

SCOPE I. Minimum Performance Requirements/Salient Characteristics Salient Characteristics The contractor shall provide all resources necessary to accomplish the tasks and deliverables described in this statement of work, including the hardware system requirements which are the minimum requirements. Equivalent requirements that differ from these minimum requirements must be justified by the proposing vendor and evaluated by the NFCC prior to purchase. 1.0 General Requirement The National Forensic Chemistry Center (NFCC) requires a Bruker Alicona InfiniteFocus G6 (IFM G6) optical 3D measurement system. This non-contact, optical 3D imaging and measurement system will be used by NFCC's Trace Examination Section (TES) for surface profilometry and difference analysis measurements, including tablet debossing, punch face embossing, surface abrasion, pit measurements, corroded surface analysis, and pharmaceutical packaging topography and print profiling. The system must expand analytical throughput to meet increasing demand for forensic 3D image analysis from law enforcement stakeholders. 2.0 Core Technology and Measurement Principle 2.1 The system shall be based on the principle of Focus Variation Microscopy (FVM), utilizing Bruker Alicona's proprietary Focus Variation algorithms to reconstruct high-resolution 3D surfaces. 2.2 The system must include advanced illumination technology (e.g., ring light, coaxial light) to enable accurate measurement of surfaces with steep angles, up to and exceeding 85 degrees. 2.3 The system must be capable of measuring both form and surface roughness parameters in a single measurement, with all data being co-registered. 2.4 The system must have the ability to measure highly reflective materials without surface treatment. The system must be capable of acquiring accurate 3D topographic data on reflective and specular materials without requiring surface alteration that could compromise the integrity of forensic samples. 2.5 The system must have the capability to capture true-color optical data co-registered with 3D topography. This capability supports surface characterization, documentation, and interpretation of coating or material variations. Many profilometry systems such as systems that use laser or tip technologies do not provide integrated true-color 3D datasets.

## Attachments

- [75F40126Q00383_final.docx](https://sam.gov/api/prod/opps/v3/opportunities/resources/files/87655dbb792849e58be4b410ab889a18/download)

## Award on USAspending

- Recipient: Bruker Nano, Inc. (UEI HF8BQ45QKKJ3)
- Contract: 75F40126P00042, purchase order
- Obligated: $323,085.00
- Competition: Competed Under SAP, 2 offers received
- Link: solicitation number 75F40126Q00383 equals the FPDS solicitation identifier; same awarding office 75F401 (high confidence)
- Record: https://www.usaspending.gov/award/CONT_AWD_75F40126P00042_7524_-NONE-_-NONE-/


## Open now in the same industry

47 open in NAICS 334519.

- STARTER,MOTOR (SPE7L327T2652), due November 27, 2026. https://sealedbidder.com/opportunities/spe7l327t2652
- REFERENCE STANDARD,PHY (SPE4A627U0220), due November 27, 2026. https://sealedbidder.com/opportunities/spe4a627u0220
- CALIBRATOR SET,RADIAC (SPE7M127T2663), due November 27, 2026. https://sealedbidder.com/opportunities/spe7m127t2663

## Publications

- September 8, 2026: Combined synopsis and solicitation, due September 14, 2026 at 3:00 AM EDT. Notice 44430997898d4b699da81c4f883899c2. https://sam.gov/workspace/contract/opp/44430997898d4b699da81c4f883899c2/view

## Points of contact

- Pamela T. Lee, pamela.lee@fda.hhs.gov, 2404023847
- Madeline Bryant, madeline.bryant@fda.hhs.gov, 2404027611

---
Source: SAM.gov Contract Opportunities bulk extract and USAspending.gov award data. Confirm deadlines on SAM.gov before responding. Cite https://sealedbidder.com/opportunities/75f40126q00383.
