# Cleanroom Testing - Sources Sought Notice Minneapolis VA Medical Center

Canonical: https://sealedbidder.com/opportunities/36c26326q0452

- Solicitation number: 36C26326Q0452
- Notice type: Sources sought
- Status: Closed. Deadline was March 24, 2026 at 6:00 PM EDT
- Department: Department of Veterans Affairs
- Contracting office: Network Contract Office 23 (36C263)
- NAICS: 541380 Testing Laboratories and Services
- Product or service code: Q301 Reference Laboratory Testing
- Small business size standard: $19 million in average annual receipts (13 CFR 121.201, NAICS 541380)
- Place of performance: Minneapolis, Minnesota
- County: Hennepin County (FIPS 27053). https://sealedbidder.com/counties/hennepin-county-mn-27053
- City: Minneapolis. https://sealedbidder.com/cities/minneapolis-mn-2743000
- First posted: March 10, 2026
- Last posted: March 10, 2026
- SAM.gov: https://sam.gov/workspace/contract/opp/f6b28b611bd84c10a20ed62094951fed/view

## Reading this notice

- Kind of notice: Sources sought. Market research. The agency wants to know who could do the work before it decides how to compete it.
- Who may compete: SAM.gov lists no set-aside on this notice, so it is not reserved for small businesses. Small, under NAICS 541380 (testing laboratories and services), means no more than $19 million in average annual receipts, counted with affiliates.
- What comes next: No solicitation was published under this number after the sources sought of March 10, 2026. An agency sometimes issues the solicitation under a new number.

## Description

The Minneapolis VA Medical Center requires Cleanroom Testing services in accordance with the draft Statement of Work (SOW) below.
This is a sources sought/proof of capability solicitation looking for vendors to perform the services per the SOW for market research purposes.

NO AWARDS OF A CONTRACT WILL BE MADE FROM THIS ANNOUNCEMENT.

If you are a vendor that can complete the work described in the SOW with competitive pricing, please send your contact and organization information with a descriptive proof of capability to: joseph.bloomer@va.gov on or before March 24th, 2026.
Only emailed responses will be considered.
Additionally, please provide answers as appropriate to the following questions in the table below with your response to this sources sought. Failure to respond to the following questions may affect the acquisition strategy.
1.
Identify your organization's socio-economic category.
2.
State whether any of the requested services may be ordered against a government contract awarded to your organization (e.g Federal Supply Schedule (FSS), General Services Administration (GSA), etc.).
3.
State if subcontracting is contemplated for this requirement and what percentage of the work will be subcontracted and for what tasks.
Â 
*If applicable, VAAR 852.219-75 VA Notice of Limitations on Subcontracting Certificate of Compliance for Services and Construction, will apply to the potential solicitation if set-aside for Veteran Owned Small-Businesses or Service Disabled Veteran Owned Small-Businesses.*
*If applicable, FAR 52.219-14 Limitations on Subcontracting, will apply to the potential solicitation if set-aside for Small-Businesses.*

Statement of Work for Environmental Microbiology Laboratory Analytical Services
GENERAL REQUIREMENTS:

An ISO/TEC 17025 accredited laboratory under the Environmental Microbiology Lab Accreditation Program (EMLAP) will process (receive, incubate, analyze, dispose, record and report results electronically) for microbiological surface and air samples, Gloved Fingertip and Thumb (GFT) samples, and Media Fill Testing (MFT) samples that will be collected and submitted by the Minneapolis VA Health Care System (MVAHCS). Contractor must be compliant will all applicable federal and local licensing boards. Samples will be processed according to the most current version of CAG-009-2023 (or its successor) and results will be compared to recommendations provided in USP  (most current version).

Sterile irradiated trypticase soy agar (TSA) plates supplemented with neutralizing additives (e.g., lecithin ad polysorbate 80) as this agar supports both bacterial and fungal growth will be provided by the contractor free of charge with no limit on the number of plates pursuant to this contract including cost of shipping the plates to the MVAHCS. The MVAHCS will cover costs of shipping the plates back to the laboratory for incubation and analysis after sample collection is completed.  MVAHCS require 2 sizes of TSA plates:
100 mm concave plates for gloved fingertip samples
55 mm convex plates for viable air and surface sampling

A certificate of analysis (COA) from the TSA plate manufacturer will be provided for each lot of growth media to verify that the media meets the expected growth promotion, pH, and sterilization requirements or the COA will be readily available online to download from the TSA plate manufacturer.

The contractor shall process personnel related assessment testing GFTs samples according to CAG-009-2023 (or its successor), and USP  2023 standards. GFT results will be reported for each plate after the initial incubation period and after the final incubation period with the amount of colony forming units (cfu) present as cfu per plate reported. If there is no growth present that shall be indicated on the report (0 CFU or  2023 standards. MFT results will be reported as Pass (negative) or Fail (positive).  Failure is indicated by visible turbidity or other visual manifestations of growth in the media in one or more container closure unit(s) on or before 14 days.

The contractor shall process viable surface and air samples according to the most current version of  CAG-009-2023 and USP  2023 standards. If cfu measured during viable sampling exceed USP  threshold levels, the report will at minimum identify genus and, when possible, species of any microorganism recovered on the report, in addition to the sample name and the number of CFU after the initial incubation period and after the final incubation period. If there is no growth present that shall be indicated on the report (0 CFU or  Table 8. If levels exceed the action levels in USP  Table 8, an attempt must be made to identify any microorganism recovered to the genus level, and when possible, to species of the microorganism recovered. A photograph of the failed plate(s) will be sent via email to the Pharmacy Program Manager for CSPs & HDs and/or designee(s).
Table 8. Action Levels for Surface Sampling
ISO Class
Surface Sampling Action Levels
(cfu / media device)
5
>3
7
>5
8
>50
USP  2023
Failure is indicated if cfu of any air sample plate is greater than the action level for the classified area based on USP  Table 7. If levels exceed the action levels in USP  Table 7, an attempt must be made to identify any microorganism recovered to the genus level, and when possible, to species of the microorganism recovered. A photograph of the failed plate(s) will be sent via email to the Pharmacy Program Manager for CSPs & HDs and  designee(s).
Table 7. Action Levels for Viable Airborne Particle Air Sampling
ISO Class
Air Sampling Action Levels
[cfu/cubic meter (1000 L) of air / media device]
5
> 1
7
> 10
8
> 100
USP  2023
Contractor will report the number of cfu on each plate after initial and final incubation as cfu per plate, with sample location and sample date. Also, the contractor will report the starting temperature for each interval of incubation, dates of incubation, and the identification of the person who reads and documents the results. As appropriate, the report will identify any microorganism exceeding the action levels as outline in USP . Contractor personnel who analyze sampling plates including reading of plates, counting cfu, and performing identification of recovered microorganisms will be a licensed and qualified environmental microbiologist.
All results will be transmitted to the facility within one business day of final analysis electronically via email to the Pharmacy Program Manager for CSPs & HDs and designee(s).

Gloved Fingertip Sampling Plates:
Pharmacy sterile compounding staff must be evaluated in Gloved Fingertip and Thumb Sampling (GFTs) every 6 months to demonstrate competence in garbing and hand hygiene, as set forth in USP . Pharmacists with direct oversight of sterile compounding must be evaluated annually in Gloved Fingertip and Thumb Sampling (GFTs) to demonstrate competence in garbing and hand hygiene, as set forth in USP . Any newly hired staff will be tested on an ad hoc basis.  MVAHCS completes competency assessments in both non-hazardous and hazardous drug sterile compounding.
Number of GFT samples per year: ~ 900  (100 mm TSA plates)
New employees:  Minimum of 30 (100 mm TSA plates)
3 sets (right and left hands) TSA plates for initial GFTs x 5 new employees per year

Technicians: Minimum of 400 (100 mm TSA plates)
25 technicians non-hazardous x 4 plates bi-annually = 200 plates
25 technicians hazardous x4 plates bi-annually = 200 plates
c. Pharmacists with direct oversight: Minimum of 280 (100mm TSA plates)
35 pharmacist non-hazardous x 4 plates annually = 140 plates
35 pharmacist hazardous x 4 plates annnually = 140 plates

d. Additional plates for re-testing:  190 (100 mm plates)
MVAHCS will collect  GFTs samples using one TSA plate per hand using sterile media containing general microbial growth agar {e.g., trypticase soy agar (TSA)} supplemented with neutralizing additives (e.g., lecithin and polysorbate 80) as this agar supports both bacterial and fungal growth.
Contractor will incubate fingertip sampling plates immediately upon receipt at a temperature of 30-35Â° C for no less than 48 hours and then at 20Â° 25Â° C for no less than 5 additional days. Sampling plates will be stored during incubation to prevent condensate from dropping onto the agar and affecting the accuracy of the cfu reading (e.g., invert plates). Contractor will record and report the number of cfu per hand (left hand, right hand) observed after initial incubation and total number of colonies observed after final incubation, with individuals name and date of testing. Also, the contractor will report the starting temperature for each interval of incubation, dates of incubation, and the identification of the person who reads and documents the results for each employee. Contractor personnel who analyze sampling plates including reading of plates, counting cfu, and performing identification of recovered microorganisms will be a licensed and qualified environmental microbiologist.

Failure is if total cfu for both hands is greater than the action level for the type sampling (e.g., initial, bi-annual, or annual). The contractor will notify the USP Program Manager, or if unavailable, the Chief Pharmacy Services, within 1 (one) business day of any failure. Contractor will provide photograph of the failed plate(s) via email to the Pharmacy Program Manager for CSPs & HDs and designee(s).

All results will be transmitted to the facility within one business day of final analysis electronically via email to the Pharmacy Program Manager for CSPs & HDs and designee(s).

Table 1. Action Level for Gloved Fingertip Sampling and Thumb Sampling

Gloved Fingertip and Thumb Sampling
Action Levels
(cfu, total from BOTH hands)
Initial after garbing x 3
> 0
On-going after garbing
> 0
After media fill
> 3
USP  2023

Media Fill Samples:
MVAHCS will collect media fill tests (MFTs) designed and developed by the MVAHCS Pharmacy Designated Person(s).  The contractor will be provided with the final MFT sample (one sample) per staff member tested.  The MVAHCS will purchase MFT kits separately outside of this contract.
Number of MFT samples per year: ~ 200  bags
a. New employees:  Minimum of 10 bags
5 new employees: 1 bag bi-annually = 10 bags

b. Technicians: Minimum of 100 bags
1) 25 technicians non-hazardous x 1 bag bi-annually = 50 bags
2) 25 technicians hazardous x 1 bag bi-annually = 50 bags

c. Pharmacists with direct oversight: 70 bags
    1) 35 pharmacist non-hazardous x 1 bag annually = 35 bags
    2) 35 pharmacist hazardous x 1 bag annnually = 35 bags
d. Additional bags for retesting: 20  bags

Contractor will incubate final MFT containers in an incubator immediately upon receipt at a temperature of 20Â° 25Â°C for 7 (seven) days followed by 7 (seven) days at 30Â° 35Â°C to detect a broad spectrum of microorganisms.
Contractor will report results as Pass / Faill as well as the starting temperature for each interval of incubation, dates of incubation, and the identification of the person who reads and documents the results. Contractor personnel who analyze the MFTs including reading results will be a licensed and qualified environmental microbiologist and/or technician.
Failure is indicated by visible turbidity or other visual manifestations of growth in the media in one or more container closure unit(s) on or before 14 days. The contractor will notify the Pharmacy Program Manager for CSPs & HDs and designee(s), within 1 (one) business day of any failure.
All results will be transmitted to the facility within one business day of final analysis electronically via email to the Pharmacy Program Manager for CSPs & HDs and designee(s).
Place of Performance:

Minneapolis VA Health Care System (MVAHCS)
Inpatient Pharmacy Department
1 Veterans Drive
Minneapolis, MN 55417
612-725-2040

Period of Performance:
Performance Years
2026 Base
9-1-2026 to 8-31-2027
2027 (Option year 1)
9-1-2027 to 8-31-2028
2028 (Option year 2)
9-1-2028 to 8-31-2029
2029 (Option year 3)
9-1-2029 to 8-31-2030
2030 (Option year 4)
9-1-2030 to 8-31-2031

Contractor Qualifications:

An ISO/TEC 17025 accredited laboratory under the Environmental Microbiology Lab Accreditation Program (EMLAP). Contractor is responsible for having the the appropriate federal, state, and local licenses. Copies of pertinent certifications, licenses, and EMLAP accreditation will be required.

Performance Requirement Summary:
Facility surface sampling analysis for viable particles from samples taken in the Segregated Compounding Area (SCA), Anteroom, Non-hazardous buffer room, hazardous drug buffer room, OR Satellite (anteroom & buffer room), Segregated Compounding Areas (3F-117A and the hazardous drug storage room), and associated Primary Engineering Controls (PECs)
Viable surface sampling analysis
Viable air sampling analysis
Identification to at least genus for any growth that exceed the levels in USP  2023 Table 7 and Table 8, and when possible, to species of the microorganism recovered.
Personnel-related competency assessments include the following:
Gloved Fingertip and Thumb Sample incubation and sample analysis
Media Fill Testing incubation and sample analysis
Surface sample of direct compounding incubation and analysis
Identification to at least genus for any growth that exceed the levels in USP  and when possible, to species of the microorganism recovered for GFTs and direct compounding area surface samples only.

Sampling Reports:
Results of viable surface and air samples, GFTs, and MFTs must be submitted to the Pharmacy Program Manager for CSPs & HDs and/or designee(s), within one (1) business days of the receipt of the results. Send reports to:
Mara Moresette, PharmD, Pharmacy Program Manager for CPS and HDs
Mara.Moresette@va.gov
Sheila Kuehne, PharmD, Associate Chief of Inpatient Operations
Sheila.Kuehne10@va.gov
Murray Leraas, PharmD, Inpatient Pharmacy Operations Supervisor
Murray.Leraas2@va.gov
Jeanna Montero, CPhT, IV Technician Supervisor
Jeanna.Montero@va.gov

a. Viable surface samples will be reported out by room location (e.g., BR-118, BR-118B,      BR-118C, BR-119, 2R-103 A & B, 3F-117A) by the contractor to the site.  Multiple rooms should not be included on the same report.

b. Viable air samples will be reported out by room location (e.g., BR-118, BR-118B, BR-118C, BR-119, 2R-103 A & B, 3F-117A) by the contractor to the site.  Multiple rooms should not be included on the same report.

If viable air and surface samples are submitted on the same chain of custody form, it will be acceptable to report both air and surface samples on the same result report for the room.

c. Each employee will have their own individual results reported by the contractor to the site.  The report will include the employee s name and clearly identify the type of test (e.g., post garbing GFT, post media fill GFT, surface sample of the DCA, and/or media fill bag). Multiple employees should not be included on the same report. Pass/Fail notifications should be included where applicable.

d. Final reports for viable air and surface sampling, GFTs and MFTs must contain the following information 100% of the time:
Name of report (e.g., name of employee or room)
Date of sampling
Date received
Date Analyzed
Date reported
Sample ID & Location
Incubation temperatures and timeframes
Media lot and expiration dates
ISO Classification
Preliminary cfu counts after initial incubation
Final cfu counts after final incubation
Total number of cfu per plate
Note if sample result is above or below USP  action level
Identification of organisms to the genus level and when possible, to species of the microorganism recovered
 Signed by lab personnel conducting the lab results.

Failure to follow these procedures may result in delay of payment or nonpayment.

Communication:
The contractor shall provide the MVAHCS Inpatient Pharmacy Service with a reliable mailing address for sample drop off within the designated business hours.
The contractor shall have ability to accept samples during usual business hours Monday to Friday excluding Federal Holidays.
The contractor shall maintain two-way communication between the Pharmacy Program Manager for CSPs and HDs and designee(s) via phone and/or email during normal business hours.
The contractor shall immediately (within 24 hours) notify the POC via phone or email upon receipt of sample(s) damaged during transit at no additional cost to the MVAHCS.
Contractor shall notify the Pharmacy Program Manager for CSPs & HDs and designee(s) of any unusual events, including but not limited to, incubator temperature excursions and other such incidents that may affect the integrity of the sample(s) within 24 hours of observation with no additional cost to the Government.
Contractor shall provide standard operating procedure on request that describes the process for handling, incubation, and analysis for each sampling plate or MFT.
Points of Contact (POC):
Mara Moresette, PharmD, Pharmacy Program Manager for CPS and HDs
Mara.Moresette@va.gov
612-467-2468
Sheila Kuehne, PharmD, Associate Chief of Inpatient Operations
Sheila.Kuehne10@va.gov
612-467-3103
Murray Leraas, PharmD, Inpatient Pharmacy Operations Supervisor
Murray.Leraas2@va.gov
612-629-7508
VA Mobile Phone: 612-986-6364

Jeanna Montero, CPhT, IV Technician Supervisor
Jeanna.Montero@va.gov
612-467-6889
Lynn Peterka, Administrative Officer / Contracting Officer Representative
Lynn.Peterka@va.gov
612-629-7929

Contingency Service:
The contractor shall immediately (within 6 hours) contact the Pharmacy Program Manager for CSPs & HDs and designee(s) via a phone call or email to the provided distrubution group of any service interruptions that may arise once identified [e.g., not passing fungal accrediation (EMPAT), etc.].  It is the responsibility of the contractor to provide contingency testing services to minimize any delay or inconveniences.
Inspection:
The Government reserves the right to inspect the contractor's facility, licenses, certificates, and equipment or require documentation of compliance.  If conducted, such inspections do not constitute a warranty by the Government that the contractor s facility, licenes, certificates, and/or equipment are properly maintained.

## Attachments

- [36C26326Q0452.docx](https://sam.gov/api/prod/opps/v3/opportunities/resources/files/7c88788cb1fd41028e92211f8c8b2400/download)

## Open now in the same industry

34 open in NAICS 541380.

- Storage and monitoring of ManNac for FDA regulations (75N98027Q00086), due November 26, 2026 at 9:00 AM EST. https://sealedbidder.com/opportunities/75n98027q00086
- APT Generator Switchgear Testing – Alvin C. York VA Medical Center (36C249-27-AP-1027), due November 26, 2026 at 1:00 PM EST. https://sealedbidder.com/opportunities/36c24927ap1027
- FY27 AASHTO ACCREDITATION AND INSPECTION SERVICES (693C7327Q000004), due November 27, 2026 at 12:00 PM EST. https://sealedbidder.com/opportunities/693c7327q000004

## Publications

- March 10, 2026: Sources sought, due March 24, 2026 at 6:00 PM EDT. Notice f6b28b611bd84c10a20ed62094951fed. https://sam.gov/workspace/contract/opp/f6b28b611bd84c10a20ed62094951fed/view

## Points of contact

- Joey Bloomer, Contracting Officer, joseph.bloomer@va.gov, 515-699-5880

---
Source: SAM.gov Contract Opportunities bulk extract. Confirm deadlines on SAM.gov before responding. Cite https://sealedbidder.com/opportunities/36c26326q0452.