# TUC/678_TWITCHVIEW 2 - NERVE SYST

Canonical: https://sealedbidder.com/opportunities/36c26226q0912

- Solicitation number: 36C26226Q0912
- Notice type: Sources sought
- Status: Closed. Deadline was June 16, 2026 at 9:00 AM EDT
- Department: Department of Veterans Affairs
- Contracting office: 262-Network Contract Office 22 (36C262)
- NAICS: 334510 Electromedical and Electrotherapeutic Apparatus Manufacturing
- Product or service code: 6515 Medical and Surgical Instruments, Equipment, and Supplies
- Small business size standard: 1,250 employees (13 CFR 121.201, NAICS 334510)
- Place of performance: Tucson, AZ, United States
- Who can bid, by location: Anywhere in the US. Location does not limit who can bid: a product shipped to the buyer, research at the performer's facility, or work at the contractor's site. A state shown is the delivery point.
- Location read from: the notice's place of performance
- County: Pima County (FIPS 04019). https://sealedbidder.com/counties/pima-county-az-04019
- City: Tucson. https://sealedbidder.com/cities/tucson-az-0477000
- First posted: June 5, 2026
- Last posted: June 5, 2026
- SAM.gov: https://sam.gov/workspace/contract/opp/6e9542f1e6db48b3945149e0746889d0/view

## Reading this notice

- Kind of notice: Sources sought. Market research. The agency wants to know who could do the work before it decides how to compete it.
- Who may compete: SAM.gov lists no set-aside on this notice, so it is not reserved for small businesses. Small, under NAICS 334510 (electromedical and electrotherapeutic apparatus manufacturing), means no more than 1,250 employees, counted with affiliates.
- What comes next: No solicitation has been published under this number yet. The sources sought of June 5, 2026 is the earliest notice of the purchase.

## Description

Sources Sought Synopsis Only

THIS REQUEST FOR INFORMATION (RFI)/SOURCES SOUGHT/RFI IS ISSUED SOLELY FOR INFORMATION AND PLANNING PURPOSES ONLY. THIS IS NOT A SOLICITATION.

THE SUBMISSION OF PRICING, CAPABILITIES FOR PLANNING PURPOSES, AND OTHER MARKET INFORMATION IS HIGHLY ENCOURAGED AND ALLOWED UNDER THIS RFI IN ACCORDANCE WITH (IAW) FAR 15.201(e).

The purpose of this synopsis is to gain knowledge of potential qualified sources and their size classification (hubzone, 8(a), small, small disadvantage, veteran owned small business (VOSB), service-disabled veteran owned small business (SDVOSB), or large business relative to NAICS 334510 with size standards of 1250 employees.  Note that if VA s pending market research establishes that two or more VOSB or SDVOSB, who are certified in U.S Small Business Administration site https://www.sba.gov can perform the project, VA will set aside the underlying solicitation per 38 U.S.C. Â§8127.Â 

Responses to this synopsis will be used by the Government to make appropriate acquisition decisions.  After review of the responses to this Sources Sought Synopsis, a solicitation announcement may be published on SAM, and/or GSA eBuy.  Responses to this Sources Sought Synopsis are not considered adequate responses to the solicitation announcement.

General Requirements:

The Department of Veterans Affairs (VA), VA Tucson Healthcare System, has a need for Twitchview 2 Monitor System and accessories. The TwitchView 2 Monitor System will be used to assess the degree of neuromuscular blockade/muscle relaxation in patients during surgery.

The requirement is for brand name or equal: TwitchView 2 System
Identify any domestic alternatives (meaning manufactured in the United States)

Nomenclature: Quantitative Neuromuscular Monitor/ EMG Monitor

MANUFACTURE AND BRAND NAME PRODUCT INFORMATION   BLINK DEVICE COMPANY

CLIN
Description
Qty
Unit of Measure
01
TwitchView 2 System w/ Pole Mount
11
EA
02
TwitchView 2 Phillips XLR Interface Cable
11
EA
03
TwitchView Adult Electrode Array (Case of 25)
1
EA
04
In-service 1-day
1
EA
05
TwitchView System Tester
1
EA

REQUIREMENTS: The Anesthesiology department requests the acquisition of the TwitchView 2 System for the purpose of tracking quantitative measurements  of a patient s motor response during operation, in post-anesthesia care units and or intensive care units. The Salient Characteristics must include:

Quantitative neuromuscular transmission monitoring system using electromyography (EMG) technology.
Must provide objective measurement of neuromuscular blockade and recovery.
Must support Train of Four (TOF) monitoring and Post Tetanic Count (PTC) monitoring.
Must include Auto PTC mode, automatically switching between TOF and PTC monitoring as needed.
Must provide real time waveform display and numerical neuromuscular parameters.
Device must use EMG waveform analysis with noise cancellation technology capable of detecting twitch responses as low as ~0.2 mV.
Must provide performance validated against mechanomyography (MMG) or equivalent clinical gold standard.
Must reliably identify TOFR   0.90 to confirm adequate recovery.
Must be designed for operating room environments with ruggedized housing.
Must include durable patient cable rated for  5,000 use cycles.
System must include a sterile, single use electrode per patient.
System must include a universal OR mounting arm compatible with common anesthesia machine rails or equipment poles.
Must include multiple electrode sizes such as:
Adult
Electrodes must support placement on hand or foot, including scenarios with tucked arms.
System must allow single button operation ( Start ) for initiating neuromuscular monitoring.
Must automatically display TOF ratio, PTC count, and neuromuscular blockade trend data.
Must maintain continuous neuromuscular monitoring throughout the procedure.
Must support electronic medical record (EMR) integration, including but not limited to:
Epic
Cerner
Innovian
Must export neuromuscular monitoring data for clinical documentation without manual transcription.
Device must have FDA 510(k) clearance for quantitative neuromuscular monitoring.
Must comply with medical electrical equipment standards applicable to OR environments.
Must include visual and/or audible alerts for electrode issues or signal loss.
Must operate using standard medical grade power supply meeting hospital electrical safety requirements.
System hardware must include a minimum 5 year warranty.
Vendor must provide technical support and replacement component as required

Participating Healthcare Facilities:

VISN 22 West
VA TUCSON Healthcare System   3601 SOUTH 6TH AVENUE TUCSON AZ; 85723

In response to this announcement, please complete the information below:

Company:
 Address:

Contact Name:
Phone No:
Email:
DUNS:

Business size information
Select all that applies:

  Small Business
  Emerging Small Business
  Small Disadvantaged Business
  Certified under Section 8(a) of the Small Business Act
  HubZone
  Woman Owned
  Certified Service-Disabled Veteran Owned Small in Vetbiz.gov
  Certified Veteran Owned Small Business in Vetbiz.gov
  Large Business
  Other:____________

All offered items must be listed on vendor FSS contract

FSS/ GSA Contract Holder of the requested item(s)
             Yes                 No
FSS/ GSA Contract Number

Effective Date / Expiration Date

Does your company manufacture the requested product(s)? If no, answer the questions below.
             Yes                 No
Does your company exceed the 500-employee alternative size standard for nonmanufacturers?
             Yes                 No
Is your company primarily engaged in the retail or wholesale trade and normally sell the type of product(s) being supplied?
             Yes                 No
Does your company take ownership or possession of the products(s) with its personnel, equipment, or facilities in a manner consistent with industry practice?
             Yes                 No
Does your company supply the end product(s) of a small business manufacturer or processor made in the US, or obtain a waiver of such requirement?
             Yes                 No

Identify any other companies owned (wholly or in-part) by the owners of this business who provide goods or services under the same or a related NAICS codes.
Not applicable
List:

Buy American Act Information
Are all items manufactured in the U.S? If not, list the items not manufactured in the US and the country of origin in the table provided below.
             Yes                 No

Item #
Description

Country of Origin

                                        Domestic Alternative Product
Do you have domestic alternatives? (Equal product that is manufactured within the United States) If so, please state the item, description and place of manufacture (country of origin) in the spaces below
             Yes                 No

Item #
Description

Country of Origin

This notice is neither a request for competitive proposal or solicitation of offerors. This notice is to assist the VA in determining sources only and a solicitation is not currently available.  If a solicitation is issued it will be announced at a later date, and all interested parties must respond to that solicitation announcement separately from this response. Responses to this Sources Sought is not a request to be added to a prospective bidders list or to receive a copy of the solicitation.  Information received as a result of this notice will normally be considered solely for the purpose of determining whether to conduct competitive procurement. This notice does not represent a commitment by the Government to pay for costs incurred in the preparation and submission of information or any other costs incurred as a response to this announcement. Inquiries will only be accepted in writing via email to Rochelle.Malveaux@va.gov on or before Tuesday, June 16th, 2026, by 6:00 am PST Local Time.

## Attachments

- [36C26226Q0912.docx](https://sam.gov/api/prod/opps/v3/opportunities/resources/files/4ae637659d61418c89f4c51787b62504/download)

## Open now in the same industry

11 open in NAICS 334510.

- NCSS - Sentec tCOM+ Transcutaneous CO2 Monitor Acq (75N98027Q00094), due November 27, 2026 at 11:00 AM EST. https://sealedbidder.com/opportunities/75n98027q00094
- Medtronic FT10 Series Cauterizer (36C26327Q0131), due November 30, 2026 at 11:00 AM EST. https://sealedbidder.com/opportunities/36c26327q0131
- FY27 Mobile CT Scanner Lease (36C24827Q0085), due November 30, 2026 at 4:00 PM EST. https://sealedbidder.com/opportunities/36c24827q0085

## Publications

- June 5, 2026: Sources sought, due June 16, 2026 at 9:00 AM EDT. Notice 6e9542f1e6db48b3945149e0746889d0. https://sam.gov/workspace/contract/opp/6e9542f1e6db48b3945149e0746889d0/view

## Points of contact

- Rochelle Malveaux, Contract Specialist, Rochelle.Malveaux@va.gov, 000-000-0000

---
Source: SAM.gov Contract Opportunities bulk extract. Confirm deadlines on SAM.gov before responding. Cite https://sealedbidder.com/opportunities/36c26226q0912.