Skip to content

SealedBidder is a private company, not a government agency.

Sign upDashboard
  • 8,711 federal solicitations are open right now. How many of them are yours?
  • New solicitations are posted every day. The ones you never saw are the ones you cannot win.
  • Stop checking SAM.gov every morning. Let the contracts that fit your business come to you.
  • Tell SealedBidder what you sell in one sentence. AI reads every open solicitation for you.
  • Deadlines are short. Hear about a solicitation the day it is posted, not the week it closes.
  • Amendments move deadlines. Watch a solicitation and SealedBidder tells you when it changes.
  • See which contracts in your field are about to end, and who holds them today.
  • Plans from $99 a month. See your matches before you pay.
See plans

Sources sought

VPRO Max2 Low Temp Sterilizers

36C25526Q0154

Department of Veterans Affairs, 255-Network Contract Office 15. Electromedical and Electrotherapeutic Apparatus Manufacturing.

Response deadline

January 21, 2026 at 12:00 PM EST

Closed 309 days ago. Posted January 14, 2026. Scheduled to archive February 5, 2026.

Get the next one like this from Department of Veterans Affairs by email, the day it is posted.Find my opportunities

This one closed in 2026

11 like it are open now in electromedical and electrotherapeutic apparatus manufacturing

  • NCSS - Sentec tCOM+ Transcutaneous CO2 Monitor Acq

    National Institutes of Health, National Institutes of Health Olao

    Combined synopsis and solicitationNAICS 334510Ship to the buyer75N98027Q00094
    Due tomorrowNov 27
  • Medtronic FT10 Series Cauterizer

    Department of Veterans Affairs, Network Contract Office 23

    Combined synopsis and solicitationSDVOSBNAICS 334510Ship to Fargo, ND36C26327Q0131
    Due in 4 daysNov 30
  • FY27 Mobile CT Scanner Lease

    Department of Veterans Affairs, 248-Network Contract Office 8

    Combined synopsis and solicitationNAICS 334510Orlando, FL36C24827Q0085
    Due in 4 daysNov 30

All 11 open in electromedical and electrotherapeutic apparatus manufacturing

Reading this notice

Worked out from the publications and the award on this page.

Kind of notice
Sources sought. Market research. The agency wants to know who could do the work before it decides how to compete it. Glossary
Who may compete
SAM.gov lists no set-aside on this notice, so it is not reserved for small businesses. Small, under NAICS 334510 (electromedical and electrotherapeutic apparatus manufacturing), means no more than 1,250 employees, counted with affiliates. What a set-aside is
What comes next
No solicitation was published under this number after the sources sought of January 14, 2026. An agency sometimes issues the solicitation under a new number.

Size standard from 13 CFR 121.201. The solicitation itself is what binds: check the code and standard it states.

Description

As published on SAM.gov.

2. SPECIFICATIONS
. Minimum of three low temperature sterilizers shall have a minimum individual dimensions of
Cabinet (DD): 33 x 40-15/16 x 72” (838 x 1040 x 1829 mm) with Recessed/Cabinet (SD): 33 x 38-
5/16 x 72” (838 x 973 x 1829 mm).
o Mounting single door, cabinet
o System installation requires no plumbing, ventilation or air
o supply – only a dedicated electrical connection is needed. A power cord is supplied for
this connection
o The facility space in which the vehicle can occupy is approx. 35’ x 110’.
. Vaporized hydrogen peroxide sterilization processes are particularly suited for sterilizing heatsensitive
materials because temperatures within the load currently do not exceed 55ºC (131ºF)
and sterilization occurs in a low-moisture environment. Heat tolerant devices and materials can
also be indicated for vaporized hydrogen peroxide sterilization processes. FDA-cleared vaporized
hydrogen peroxide sterilizers, either with or without plasma, are currently available in the
United States.
. Per ANSI/AAMI ST58:2013 or subsequent, the sterilizer shall be installed and operated in an area
with a ventilation system capable of at least 10 air exchanges per hour.
. VPRO MaX2 must support utility requirements:
o Current 16 AMP
o Minimum Circuit Breaker 20 AMPS
o Use NEMA L15-20R Plug
o Provide grouped or ganged circuit protection and disconnect for phase 3 power as
required by NEC and local codes.
o Individual power shutoffs near each machine for servicing.

Attachments

Files listed on SAM.gov. The ones we hold a copy of download from here; the rest open SAM's download, which needs no account.

Publications

Every notice SAM.gov issued under this solicitation number, oldest first. Each is a separate record on SAM.

  1. January 14, 2026

    Sources sought

    Due January 21, 2026 at 12:00 PM EST. SAM.gov, notice 4211ea88674841a19b166c31a5acd9f8

    2 revisions on SAM
    • January 26, 2026: publish, presolicitation
    • January 14, 2026: publish, sources sought

Points of contact