# MRI CONTRAST INJECTOR NAME BRAND: GUERBET OPTISTAR ELITE

Canonical: https://sealedbidder.com/opportunities/36c25525q0424

- Solicitation number: 36C25525Q0424
- Notice type: Sources sought
- Status: Closed. Deadline was June 18, 2025 at 9:00 AM EDT
- Department: Department of Veterans Affairs
- Contracting office: 255-Network Contract Office 15 (36C255)
- NAICS: 339112 Surgical and Medical Instrument Manufacturing
- Product or service code: 6515 Medical and Surgical Instruments, Equipment, and Supplies
- Small business size standard: 1,000 employees (13 CFR 121.201, NAICS 339112)
- Place of performance: Marion, Illinois
- Who can bid, by location: Anywhere in the US. Location does not limit who can bid: a product shipped to the buyer, research at the performer's facility, or work at the contractor's site. A state shown is the delivery point.
- County: Williamson County (FIPS 17199). https://sealedbidder.com/counties/williamson-county-il-17199
- City: Marion. https://sealedbidder.com/cities/marion-il-1746916
- First posted: June 11, 2025
- Last posted: June 11, 2025
- SAM.gov: https://sam.gov/workspace/contract/opp/257eee512eca48a3964422ee624f7cd9/view

## Reading this notice

- Kind of notice: Sources sought. Market research. The agency wants to know who could do the work before it decides how to compete it.
- Who may compete: SAM.gov lists no set-aside on this notice, so it is not reserved for small businesses. Small, under NAICS 339112 (surgical and medical instrument manufacturing), means no more than 1,000 employees, counted with affiliates.
- What comes next: No solicitation was published under this number after the sources sought of June 11, 2025. An agency sometimes issues the solicitation under a new number.

## Description

STATEMENT OF WORK
Description of Use:  For administering contrast injection of diagnostic Magnetic Resonance Imaging (MRI) examination in the Marion IL VA Healthcare System MRI Suite within Imaging Service.
Replacement of end-of-life cycle Magnetic Resonance (MR) dual syringe contrast media and flush solution injection system to be used in conjunction with GE Signa Artist 1.5 MRI Scanner.  Brand Name, Guerbet OptiStarÂ® Elite required for compatibility with prefilled syringes used by MRI Service.
Date of manufacture must not be greater than one (1) year before the purchase date.  The equipment must not be used or refurbished.
Training and installation must be included.
Provide Trade-In for existing Guerbet OptiStarÂ®
Must be compatible with and/or capable of :
Interfacing with GE Signa Artist 1.5 MRI Scanner
Prefilled Syringes DOTAREM contrast NDC:  67684-3001-02 and saline syringes NDC:  00019-1188-81
Minimum Technical Specification:
To be used with prefilled syringe and vials
Product will be mobile with installation of rolling pedestal design
Syringe compatibility and Flow Rate:
Side A:  60 mL syringes at 0.1   10.0 mL/sec; 10-, 15-, 20-, and 30-mL syringes at 0.1   0.8 mL/sec
Side B:  60 mL single-use disposable syringes at 0.1   0.8 mL/sec; 50- and 125- mL prefilled saline syringes* at 0.1   0.8 mL/sec (*must fit cannula down to 6mm)
Pressure Limit:
60 mL empty syringes: 150 psi
10-, 15-, 20-, and 30-mL prefilled syringes:  200 psi*
50- and 125-mL prefilled saline syringes:  150 psi*
Volume:
Side A:  10-, 15-, and 20-mL syringes:  1 mL to 20 mL; 30- and 60-mL syringes:  to installed syringe size.
Side B:  60 mL syringes:  1 mL to installed syringe size; 50- and 125-mL prefilled saline syringes*: 1 mL to installed syringe size
Scan Delay:  0-60 seconds
Inject Delay:  0-300 seconds
Number of Phases:  6 including Timing BolusÂ® feature
Number of Programs:  40
Inject History:  Last 48 injected protocols
MR Compatibility:  Up to 3 Tesla
Patency test mode option to check vein prior to injection (dedicated powerhead key)
Service contract for system maintenance with annual inspection (this service will be for a base plus four (4) option years).

Optional/Value Added Features:
IV fluid drip option for continuous IV-line patency throughout the examination.
Timing Bolus: Test injection to confirm protocol
Pressure:  Automatic sensing & setting to ensure best security
Direct power supply (e.g. not reliant on batteries)
Touchscreen console with clearly identifiable color command keys
Clear indicators & status line
Remote begin and end injection options
Programable injection protocol storage: up to 6 phases with Timing BolusÂ®, even more with injection & phase delays
Powerhead keys

## Attachments

- [36C25525Q0424.docx](https://sam.gov/api/prod/opps/v3/opportunities/resources/files/e76b495d209e456fafc488a62e992870/download)

## Open now in the same industry

15 open in NAICS 339112.

- Automated External Defibrillator (AED) (W911S227U0200), due November 26, 2026 at 3:15 PM EST. https://sealedbidder.com/opportunities/w911s227u0200
- Vivarium Tunnel Washer (36C24727Q0119), due November 30, 2026 at 8:00 AM EST. https://sealedbidder.com/opportunities/36c24727q0119
- NX EQ Microscope, Light, Operating, Hand/Plastic Surgery (VA-27-00017958) (36C10X27Q0025), due November 30, 2026 at 10:00 AM EST. https://sealedbidder.com/opportunities/36c10x27q0025

## Publications

- June 11, 2025: Sources sought, due June 18, 2025 at 9:00 AM EDT. Notice 257eee512eca48a3964422ee624f7cd9. https://sam.gov/workspace/contract/opp/257eee512eca48a3964422ee624f7cd9/view

## Points of contact

- TIMOTHY SCARBOROUGH, Contract Specialist, timothy.scarborough@va.gov, 913-946-1982

---
Source: SAM.gov Contract Opportunities bulk extract. Confirm deadlines on SAM.gov before responding. Cite https://sealedbidder.com/opportunities/36c25525q0424.