# Thyroid Uptake System, Poplar Bluff, MO VAMC

Canonical: https://sealedbidder.com/opportunities/36c25525q0328

- Solicitation number: 36C25525Q0328
- Notice type: Solicitation
- Status: Awarded to Dga Medical LLC
- Department: Department of Veterans Affairs
- Contracting office: 255-Network Contract Office 15 (36C255)
- NAICS: 334510 Electromedical and Electrotherapeutic Apparatus Manufacturing
- Product or service code: 6515 Medical and Surgical Instruments, Equipment, and Supplies
- Set-aside: Service-disabled veteran-owned small business set-aside
- Small business size standard: 1,250 employees (13 CFR 121.201, NAICS 334510)
- Place of performance: Poplar Bluff, Missouri
- Who can bid, by location: Anywhere in the US. Location does not limit who can bid: a product shipped to the buyer, research at the performer's facility, or work at the contractor's site. A state shown is the delivery point.
- County: Butler County (FIPS 29023). https://sealedbidder.com/counties/butler-county-mo-29023
- City: Poplar Bluff. https://sealedbidder.com/cities/poplar-bluff-mo-2959096
- First posted: May 9, 2025
- Last posted: May 9, 2025
- SAM.gov: https://sam.gov/workspace/contract/opp/c8aed55f45714c4ea51fe0d57b35e8cd/view

## Reading this notice

- Kind of notice: Solicitation. A formal request for proposals, quotes or bids. Offers are due by the response deadline.
- Who may compete: Service-disabled veteran-owned small business set-aside. Reserved for service-disabled veteran-owned small businesses. Small, under NAICS 334510 (electromedical and electrotherapeutic apparatus manufacturing), means no more than 1,250 employees, counted with affiliates.
- Time to respond: Bidders had 7 days from the day the solicitation was posted, May 9, 2025, to the deadline.
- How it ended: Awarded to Dga Medical LLC on June 11, 2025, 26 days after offers were due. The contracting office reported 1 offer (competed under simplified acquisition procedures).

## Description

STATEMENT OF REQUIREMENT

A. GENERAL INFORMATION

The Nuclear Medicine service at John J. Pershing VA Medical Center (VAMC) in Poplar Bluff, MO will better serve its Veterans by acquiring this Thyroid Uptake System with Well Counting System with 2 Inch Well Shielding, a multi-purpose spectrum analysis instrument designed for diverse nuclear medicine applications. The system performs Uptake studies, wipe tests, and other user-defined tasks that are accomplished with speed and precision using this fully integrated computer-controlled instrument and its comprehensive selection of application software.

Engineered for mobility, durability and operational efficiency, the unit handles clinical tests, safety compliance tasks, and system administration procedures quickly and accurately. Built-in menus follow conventionally established nuclear medicine procedures, providing step-by-step guidance throughout defined procedures. The system automatically performs calculations, stores patient information and test results, and outputs clear, concise reports. User-defined uptake protocols can be initiated.

B. SCOPE

Delivery of the Thyroid Uptake system will allow a bioassay program for quick and efficient measurement of the staff thyroid burden for radiopharmaceuticals, I-123, I-125 and I-131. Measurement results can be reported in cpm, dpm, uCi, or Bq. Action levels may be set by the user for each isotope. System reporting facilitates single bioassay report, summary of multiple bioassays on an individual staff member, and summary report on multiple staff members. Reports are easily printed or saved as PDF files

The Thyroid Uptake program supports multiple time-stamped uptake measurements, auto decay correction, user-defined uptake protocols, on-screen spectrum acquisition and analysis. The system provides reports that include normal ranges and allow notes/comments entered by the facility physician and technologists.

The wipe test application automatically calculates net contamination levels and reports results in dpm, uCi and Bq. The program allows compliance with regulations for counting and reporting results of wipes taken for contamination surveys. The system allows user to define an unlimited number of &ldquo;Area&rdquo; names and wipe &ldquo;Locations&rdquo; entered within the areas for testing and documentation. Each area may be designated as &ldquo;restricted&rdquo;, &ldquo;unrestricted&rdquo;, &ldquo;package&rdquo; or &ldquo;sealed source&rdquo;. The user may specify count times, activity units, wide window, and individual isotopes and geometric efficiencies or simply utilize the wide range of factory default options.

The program displays and stores the full spectrum with each wipe, automatically calculates net contamination levels and reports results in dpm, uCi or Bq. Wipes above the user defined trigger levels are easily identified. A detailed spectrum analysis may

also be performed for any wipe to help identify isotopes causing contamination. Customized reports can be printed or saved in PDF format.

The application specialist will provide training and guidance to ensure safe use of the equipment and the study and administration applications provided. The contractor will provide cybersecurity best practices information and training, when applicable.

The contractor will provide application update to ensure information security in the current equipment and other factors that interconnects with the imaging system including the network, other current peripheral applications and other new tools and software that will be developed or installed to work with the current work station and imaging equipment hardware and software.

C. REQUIREMENT/SALIENT CHARACTERISTICS/SPECIFICATIONS

Common Nomenclature: Medical Spectrometer Hardware

Kind of material: ALL IN ONE Touch Screen Computer

Electrical Data:

Power: Low voltage transformer with hospital grade power cord and plug.

Line Voltage: 120 VAC, 60 Hz, 2 AMP circuit breaker; 230 VAC, 50 Hz, 1 AMP circuit breaker

Dimensions, size, capacity:

Minimum articulating arm with a minimum of 34-inch vertical range of travel with a minimum of 220-degree collimator rotation.

Principles of operation:

Multi-Channel Analyzer:

Channels: 1024

Spectral Resolution: FWHM 8%

Count Rate: (Maximum) 150,000 cps for Tc-99m

Count Rate Stability: 99%

Gross Count Rate Linearity: Within 5% up to 150,000 cps

Pulse Height Linearity: Within 2% (independent of detector)

Detector High Voltage Adjustment: Automatic adjustment for both probe and well; uses 10 &mu;Ci Cs-137 as the calibration source.

Restrictive environmental conditions: User Set: Unlimited user defined Isotopes, setting ROI, half-life, name, efficiency and range.

Intended use:

Factory Programmed: Au-198, Ba-133, Co-57, Co-58, Co-60, Cr-51, Cs-137, F-18, Fe-59, Ga-67, Ga-68, Gd-153, Ge-68/Ga-68, Hg-197, I-123, I-125, I-131, In-111, Ir-192, K-42, Lu-177, Na-22, Na-24, Pd-103, Ra-223, Se-75, Sr-85, Tc-99m, Tc-99m/Tl-201, Tl-201, Yb-169, Wide Window.

Equipment with which the item is used, Job related tools:

Probe: 2" x 2" Nal (Tl) integral line scintillation detector

Other pertinent information that describes the item, material or service required:

Uptake Stand:

Dimensions:

Arm down and retracted: 34" l x 26" w x 55" h (86.4 x 66 x 139.7 cm)

Maximum Height: 60" (152.4 cm)

Maximum Length: Arm Extended: 61" (154.9 cm)

Collimated Shield: Flat field collimator meeting IAEA specifications

Positioning: Distance bar and positioning light

Casters: 4" Total locking

Maximum weigh of 300 lbs.

A maximum combined weight with Well Counter: 350 lbs.

Power: Low voltage transformer with hospital grade power cord and plug.

Line Voltage: 120 VAC, 60 Hz, 2 AMP circuit breaker; 230 VAC, 50 Hz, 1 AMP circuit breaker

OTHER UNIQUE REQUIREMENTS: Minimum of two-inch well container

Compatibility with Existing Equipment: TRM Compliance with VA Directives

WINDOWS 10 compatible or Newest Windows Version compliance with VHA standards.

Must be DICOM compliance, Directives or IT Policies, Required.

Coordination with On-going Construction Project (provide the project number): N/A

Warranty: 1 year from date of receipt and verification of installation/operation. Two years part/labor warranty

Training: Initial installation and launch training staff and service support

Service and/or Maintenance: Preventive maintenance and training support, to include two Operating Manuals and two Service Manuals

Safety/ISO/IT/BioMed/SPS & Other Departmental Compliance: Must meet Technical Reference Model (TRM) VA/BioMed Compliance.

D. DELIVERY

Contractor must deliver all components of the system within 30 days of award notice.

Standard FOB Destination to:

John J. Pershing VA Medical Center

1500 North Westwood Blvd

Poplar Bluff MO 63901

Government-Furnished Property, Material, Equipment, or Information (GFP, GFM, GFE, or GFI) N/A

SECURITY N/A

SPECIAL MATERIAL REQUIREMENT: Two Operator&rsquo;s Manuals and two Service Manuals

E. PLACE OF PERFORMANCE

The system will be installed, serviced, and updated at the facility, John J. Pershing VAMC, Poplar Bluff, MO, Nuclear Medicine Department. Servicing, training, and support of personnel will be per stipulations on the final contract.

## Attachments

- [36C25525Q0328.pdf](https://sam.gov/api/prod/opps/v3/opportunities/resources/files/a825ea5ba6ad4aeabd7d5d59b0f20fc5/download)

## Award on USAspending

- Recipient: Dga Medical LLC (UEI GZT6NJ8HJLC3)
- Contract: 36C25525P0332, purchase order
- Obligated: $34,933.00
- Competition: Competed Under SAP, 1 offers received
- Link: solicitation number 36C25525Q0328 equals the FPDS solicitation identifier; same awarding office 36C255 (high confidence)
- Record: https://www.usaspending.gov/award/CONT_AWD_36C25525P0332_3600_-NONE-_-NONE-/


## Open now in the same industry

11 open in NAICS 334510.

- NCSS - Sentec tCOM+ Transcutaneous CO2 Monitor Acq (75N98027Q00094), due November 27, 2026 at 11:00 AM EST. https://sealedbidder.com/opportunities/75n98027q00094
- Medtronic FT10 Series Cauterizer (36C26327Q0131), due November 30, 2026 at 11:00 AM EST. https://sealedbidder.com/opportunities/36c26327q0131
- FY27 Mobile CT Scanner Lease (36C24827Q0085), due November 30, 2026 at 4:00 PM EST. https://sealedbidder.com/opportunities/36c24827q0085

## Publications

- May 9, 2025: Solicitation, due May 16, 2025 at 12:00 PM EDT. Notice c8aed55f45714c4ea51fe0d57b35e8cd. https://sam.gov/workspace/contract/opp/c8aed55f45714c4ea51fe0d57b35e8cd/view

## Points of contact

- Tywanna Sparks, tywanna.sparks@va.gov, 9137589928

---
Source: SAM.gov Contract Opportunities bulk extract and USAspending.gov award data. Confirm deadlines on SAM.gov before responding. Cite https://sealedbidder.com/opportunities/36c25525q0328.