Sources sought
Class II, A2 Biosafety Cabinet
36C25525Q0142
Department of Veterans Affairs, 255-Network Contract Office 15. Surgical Appliance and Supplies Manufacturing.
Response deadline
January 29, 2025 at 12:00 PM EST
Closed 667 days ago. Posted January 22, 2025. Scheduled to archive February 13, 2025.
Get the next one like this from Department of Veterans Affairs by email, the day it is posted.Find my opportunities
This one closed in 2025
20 like it are open now in surgical appliance and supplies manufacturing
- SOURCES SOUGHT NOTICE FOR HEAD, SHOULDER, AND UPPER-EXTREMITY POSITIONING ACCESSORIES FOR THE MAQUET MEERA 7200 OPERATING TABLE
Department of Veterans Affairs, 260-Network Contract Office 20
Sources soughtNAICS 339113Ship to Spokane, WA36C26027Q0130Due todayNov 27 - VALVE,BREATHING APP
Department of the Navy, NAVSUP Weapon Systems Support Mech
SolicitationNAICS 339113Ship to the buyerN0010427QLA17Due todayNov 27 - 36C257-27-AP-0953 Prosthetic STX 671 O2 Equipment (VA-27-00016458)
Department of Veterans Affairs, 257-Network Contract Office 17
SolicitationSDVOSBNAICS 339113Ship to San Antonio, TX36C25727Q0104Due in 3 daysNov 30
All 20 open in surgical appliance and supplies manufacturing
Reading this notice
Worked out from the publications and the award on this page.
- Kind of notice
- Sources sought. Market research. The agency wants to know who could do the work before it decides how to compete it. Glossary
- Who may compete
- SAM.gov lists no set-aside on this notice, so it is not reserved for small businesses. Small, under NAICS 339113 (surgical appliance and supplies manufacturing), means no more than 800 employees, counted with affiliates. What a set-aside is
- What comes next
- No solicitation was published under this number after the sources sought of January 22, 2025. An agency sometimes issues the solicitation under a new number.
Size standard from 13 CFR 121.201. The solicitation itself is what binds: check the code and standard it states.
Description
As published on SAM.gov.
Objective
The VA Saint Louis Health Care System intends to procure a Class II, Type A2 biosafety cabinet. This document highlights the technical specifications and services being requested by the facility for consideration towards purchase of the units. Vendors under this proposal shall provide all equipment and accessories, installation services, and project management support.
Equipment Specifications
Requirements
- Class II, Type A2 biosafety cabinet
- 10 inch minimum sash opening
- Two electrical duplex receptacles
- Supply & exhaust HEPA filters with 99.99% efficiency
- Noise level < 65 dBA
- NSF certified
- ADA-compliant control
- Stainless steel interior & powder coated steel exterior
- Minimum one year manufacturer warranty
System functionality
- Product must be fully installed by vendor.
Installation and Assembly
- Vendor shall be responsible for assembly and installation of systems after notification from VA Saint Louis Biomedical Engineering that it has been delivered to station.
- Installation shall include implementation and testing of all purchased interfaces to VA electronic systems in coordination with Clinical Staff, Biomedical Engineering staff, O&IT staff, and others identified as needed.
- Install shall include unboxing and transportation of equipment to final use location.
- Assembly and install shall occur within 30 days of delivery to station.
- Vendor shall provide all necessary accessories, cables, adaptors, etc. to deliver a fully functional clinical device.
Warranty and Service
- Service and Operator Manuals – The vendor shall provide the following documentation for the proposed system:
- Two (2) copies of operator's instruction manuals per unit purchased
- Two (2) copies of complete technical service manuals including detailed troubleshooting guides, necessary diagnostic software, service keys, schematic diagrams, parts lists, and any other resources available to the manufacturer’s service technicians.
- Two (2) copies of a system manager’s manual outlining back-up procedures, managing privilege group limits, routine tasks, etc.
- Warranty – The system, all accessories, and all software shall be covered under the manufacturer’s warranty and shall include all parts and labor for at least one year following acceptance by the VAMC.
- A manufacturer’s factory-trained Field Service Engineer shall perform installation and maintenance during the warranty period.
- The warranty period shall not begin until after the VA Saint Louis Biomedical Engineering POC has accepted the products delivery, installation, and functionality.
- The warranty shall include all manufacturer recommended Preventive Maintenance procedures for its duration.
- All available software updates and security patches will be made available to the VA Saint Louis during the warranty period.
- Vendor shall provide response to applications and technical calls within 8 hours.
- Vendor shall ship replacement parts within 48 hours of identifying necessity by VA Saint Louis Biomedical Engineering staff.
Hours of Operation
- Normal hours of coverage are Monday through Friday, 7:00 AM to 3:30 PM Central Standard Time excluding national holidays.
- Vendor on-site visits will be coordinated with the VA Saint Louis Biomedical Engineering POC before being scheduled or conducted.
- Vendor representatives will check in with the VA Saint Louis Biomedical Engineering POC when they arrive on-site and check out upon completion of work each day.
- All services shall be performed during normal hours unless otherwise scheduled with the VA Saint Louis Biomedical Engineering POC.
- Vendor staff shall always be professional and shall take all necessary precautions to maintain a safe environment for the occupants. This includes, as necessary, coordinating access and closures for sensitive areas of the Medical Center.
- Vendor shall be responsible for cleaning all work areas after completion of tasks. This includes removal of packaging and a clean sweep/mop of the area as required by work type.
- Vendor staff shall minimize noise as much as possible while working in all buildings.
- Vendor is responsible for providing all PPE necessary for their work.
Federal Holidays observed by VA Saint Louis are:
New Year’s Day
Labor Day
Martin Luther King Day
Columbus Day
President’s Day
Veterans’ Day
Memorial Day
Thanksgiving Day
Juneteenth
Christmas Day
Independence Day
Delivery
- Vendor shall deliver purchased equipment within 60 business days of receiving completed Contracting package from the VISN 15 Contracting Officer. If delivery delays occur vendor shall notify the Contracting Officer and the VA Saint Louis Biomedical Engineering POC.
- Vendor will ensure that equipment is sent to the Attention of Biomedical Engineering and will include the VA Saint Louis PO Number on the shipping label.
- Delivery of oversized, palletized orders or orders that contain a large number of components will be coordinated with VA Saint Louis Biomedical Engineering POC who will communicate internally with Warehouse Staff.
Attachments
Files listed on SAM.gov. The ones we hold a copy of download from here; the rest open SAM's download, which needs no account.
- Statement of Requirement.docx27 KB, January 22, 2025
- Sources Sought 36C25525Q0142.pdf194 KB, January 22, 2025
Publications
Every notice SAM.gov issued under this solicitation number, oldest first. Each is a separate record on SAM.
January 22, 2025
Sources sought
Due January 29, 2025 at 12:00 PM EST. SAM.gov, notice 90890c8b3e6a4672bc5dcfc200d335ba
2 revisions on SAM
- February 21, 2025: publish, solicitation
- January 22, 2025: publish, sources sought
Points of contact