Sources sought, service-disabled veteran-owned small business sole source
Urinalysis Analyzer & Testing Services (Base + Four Years)
36C249-24-AP-0211
Department of Veterans Affairs, 249-Network Contract Office 9. Analytical Laboratory Instrument Manufacturing.
Response deadline
January 26, 2024 at 1:00 AM EST
Closed 1035 days ago. Posted January 19, 2024, first published January 19, 2024. Scheduled to archive February 10, 2024.
Get the next one like this from Department of Veterans Affairs by email, the day it is posted.Find my opportunities
This one closed in 2024
53 like it are open now in analytical laboratory instrument manufacturing
- Benchtop biophysical instrument
Department of Energy, ORNL Ut-Battelle LLC-DOE Contractor
Combined synopsis and solicitationNAICS 334516Ship to Oak Ridge, TNRFQ519526Closes in 9 daysDec 5 - NANO-AEROSOL HYGROSCOPICITY- CLOUD CONDENSATION ANALYZER
Department of the Air Force, FA8601 AFLCMC Pzio
Sources soughtNAICS 334516Ship to Wright-Patterson AFB, OHFA860127Q0031Due tomorrowNov 27 - Notice of Intent - Illumina MiSeq i100 series instruments
National Institutes of Health, National Institutes of Health - CC
PresolicitationNAICS 334516Ship to Bethesda, MDNOI-OLAO-CC-27-000836Due tomorrowNov 27
All 53 open in analytical laboratory instrument manufacturing
Reading this notice
Worked out from the publications and the award on this page.
- Kind of notice
- Sources sought. Market research. The agency wants to know who could do the work before it decides how to compete it. Glossary
- Who may compete
- Service-disabled veteran-owned small business sole source. Awarded to one service-disabled veteran-owned small business without competition. Small, under NAICS 334516 (analytical laboratory instrument manufacturing), means no more than 1,000 employees, counted with affiliates. Open SDVOSB sole source opportunities
- What comes next
- No solicitation was published under this number after the sources sought of January 19, 2024. An agency sometimes issues the solicitation under a new number.
Size standard from 13 CFR 121.201. The solicitation itself is what binds: check the code and standard it states.
Description
As published on SAM.gov.
Location VA Knoxville Community Based Outreach Clinic (CBOC) 1557 Downtown West Boulevard Knoxville, TN 37919-5407 Performance Work Statement (PWS) – Urinalysis Analyzer & Support Services 1. Technical Features. Instrumentation shall perform all analyses with precision as defined by current Clinical Laboratory Improvement Amendments (CLIA) and the Clinical and Laboratory Standards Institute (CLSI). 2. Operational Features. Instrumentation must be available at the time of proposal submission and shall:
a. Function as an automated, walk-away instrument for urine chemistry and microscopic analysis.
b. Provide identification for the following particles in urine: RBC, WBC, Bacteria, Hyaline Casts, Pathological Casts, Crystals, Squamous Epithelial Cells, Non-Squamous Epithelial Cells, Yeast, WBC Clumps, Sperm and Mucus.
c. Provide reagent barcode capability and inventory control, to include data archiving and active reagent volume monitoring and warning. Reagents shall have extended calibration stability.
d. Read bar coded patient samples using the VistA Universal Identification system and have the ability to enter/run patient samples manually, with or without barcodes, during HIS/LIS downtime. • This analyzer connects to VISTA via a DICOM interface. • The contractor will have no access to patient information during anytime (i.e. during setup/ installation, routine maintenance, or service.
e. Minimize sample carryover: Carryover specifications shall be provided for each instrument.
f. Include a multitasking computer system with:
(1) User-friendly software. Operators of all skill levels should be able to use efficiently.
(2) Ability to locate patient samples using patient name, SSN, or unique ID, and determine when results will be available.
(3) Ability to print patient reports -in a chartable/medical record format- that may be sent to providers during HIS/LIS downtime.
(4) Ability to retransmit results after HIS/LIS downtime.
(5) Ability to easily add/update quality control and calibrator materials and values, and track the changes.
(6) Ability to run new and old lot numbers of QC concurrently for parallel testing.
(7) Ability to use only one printer to print QC data, review screens, or print files while running patient samples.
(8) Ability to review abnormal results on the instrument.
g. Include an interactive, on-board QC package with:
(1) Modified Westgaard rules and Levy-Jennings plots.
(2) Printable QC results that include meaningful information, such as Standard Deviation Index.
(3) Electronic archiving and corrective action documentation.
(4) Ability to evaluate and print QC results while the instrument is analyzing patient samples.
(5) Ability to monitor calibrations for slope, intercept, blank, etc.
(6) Ability to export values to a commercial quality control program.
h. Include bi-directional LIS host query interfacing through Data Innovations. The vendor shall provide additional equipment, such as middleware, needed to interface the analyzers. The analyzer shall be able to retransmit data after instrument downtime. The instrument should only require a single LIS Interface.
i. Provide true STAT interrupt.
j. Offer a throughput of at least 70 tests per hour.
k. Require minimal time to perform recurring maintenance (daily, weekly, monthly, etc.).
l. Require minimal start up time from standby mode to operation.
m. Require minimal reagent and calibration preparation.
n. System shall have detector system for bubbles, clots, and incomplete sample aspirations. System should be available for detecting interfering substances.
o. Include on-board electronic operating manuals and trouble shooting guide as well as a preventive maintenance guide and log.
p. Ensure ease of calibration with extended calibration stability and minimal calibration frequency.
q. System shall be totally self-contained excluding waste and will not require external water, pressure or vacuum. Vendor shall provide a waste stream analysis for each proposed instrument and if any waste requires special handling before disposal.
r. Require minimal time to perform recurring maintenance (daily, weekly, monthly, etc.). System must have a cleansing mechanism for small clots, electronic cleansing mechanism preferred.
s. Require minimal start up time from standby mode to operation.
t. System shall be calibrated on assayed material compatible with CAP guidelines. The analyzer shall compile all calculation. Cal factors shall be stored internally and activated internally at the operator’s discretion. 3. Optional Features. In addition to the base capabilities listed above, optional/desirable capabilities may be required as indicated below.
a. Touch screen.
b. Meter or computer printout for calibrators, controls, patient results, repeats.
Publications
Every notice SAM.gov issued under this solicitation number, oldest first. Each is a separate record on SAM.
Points of contact
- Mark MiezinMark.Miezin@va.gov6153472129