# Bariatric Beds and Surface Mattresses-509

Canonical: https://sealedbidder.com/opportunities/36c24725q0421

- Solicitation number: 36C24725Q0421
- Notice type: Sources sought
- Status: Closed. Deadline was April 15, 2025 at 10:00 AM EDT
- Department: Department of Veterans Affairs
- Contracting office: 247-Network Contract Office 7 (36C247)
- NAICS: 339113 Surgical Appliance and Supplies Manufacturing
- Product or service code: 6515 Medical and Surgical Instruments, Equipment, and Supplies
- Small business size standard: 800 employees (13 CFR 121.201, NAICS 339113)
- Place of performance: Augusta, GA
- Who can bid, by location: Anywhere in the US. Location does not limit who can bid: a product shipped to the buyer, research at the performer's facility, or work at the contractor's site. A state shown is the delivery point.
- First posted: April 10, 2025
- Last posted: April 10, 2025
- SAM.gov: https://sam.gov/workspace/contract/opp/ee039dcfa8f74067953baf73b8c80f29/view

## Reading this notice

- Kind of notice: Sources sought. Market research. The agency wants to know who could do the work before it decides how to compete it.
- Who may compete: SAM.gov lists no set-aside on this notice, so it is not reserved for small businesses. Small, under NAICS 339113 (surgical appliance and supplies manufacturing), means no more than 800 employees, counted with affiliates.
- What comes next: No solicitation was published under this number after the sources sought of April 10, 2025. An agency sometimes issues the solicitation under a new number.

## Description

This is a sources  sought announcement only. All responses will be used for market analysis in determining the availability of potential qualified businesses that may be used to determine appropriate acquisition strategy. No solicitation is being issued at this time but may be issued later. The Columbia Healthcare VA System (WJB Dorn VA Medical Center) located at 6439 Garners Ferry Road, Columbia, SC 29209 on behalf of the Augusta, VA Medical Center is seeking companies that are capable to provide bariatric bed frames and therapy mattresses.           Must meet or exceed all stated salient characteristics identified below.

The bariatric bed must have a patient weight range 250-1000 lbs. (113-454 kg) with a maximum weight capacity of 1000 lbs.
The bariatric bed must have a low position sleep deck to floor 18.5" (47 cm)
The bariatric bed must have a high position sleep deck to floor 26.8" (68 cm)
The bariatric bed must have a transport position sleep deck to floor 17" (43 cm)
The bariatric bed must have a minimum overall width (width extenders retracted) 43" (109 cm)
The bariatric bed must have a minimum overall length (foot extension retracted) 91" (231 cm)
The bariatric bed must have a maximum patient weight capacity 1,000 lbs (453.6 kgs)
The bariatric bed must have a powered width expansion 40" - 50" (101 or 127 cm)
The bariatric bed must have a powered length adjustment 80" - 88" (203 cm - 224 cm)
The bariatric bed must have a digital head-of-bed Indicator and Alert
The bariatric bed must have a scale
The bariatric bed must have a scale system that must have a Accuracy 1% of patient weight (with sleep deck flat), weighs in any position, and have a scale weight capacity 1,000 lbs (453.6 kgs)
The bariatric bed must have a battery backup
The bariatric bed must have a nightlight
The bariatric bed must have a integrated caregiver and patient controls in side rail
The bariatric bed must have a frame that reduces patient from sliding, i.e. reduce patient migration
The bariatric bed must have a power motorized transport drive with 6" casters
The bariatric bed must have a integrated bed exit alarm with three levels of sensitivity
The bariatric bed must have a Nurse Call ready
The bariatric bed must have a Nurse Call capability and Nurse call located in Siderail
The bariatric bed must have a removable single-post trapeze bar
The bariatric bed must have a bed controls and Nurse call button in the siderail
The bariatric bed must have a bed controls and Nurse call button in the pendant
The bariatric bed must have a CPR quick-release handle
The bariatric bed must have a English Language Labels
The bariatric bed must have an IV Pole
The bariatric bed must have a line management system at top of bed frame located at head-of-bed right and left side which holds IV lines in place
The bariatric bed must have a Caster size 6" (15 cm)
The bariatric bed must have a four-wheel braking
The bariatric bed must have a brake not set alert
The bariatric bed must have a brakes lock both rolling & swiveling
The bariatric bed must have a Head section 0-50Â°
The bariatric bed must have a Head-of-bed angle indicator
The bariatric bed must have a 30 degree Head-of-bed angle alert
The bariatric bed must have a Knee section 0-30Â°
The bariatric bed must have a Max Tilt/Reverse Tilt 0-9Â°
The bariatric bed must have a Power requirements 120v, 60 cycle AC
Fuse 15A, with a maximum current 9.9A,
The bariatric bed must have a straight power cord
The bariatric bed must have a low air loss surface with weight-based pressure redistribution surface, turn assist, and integrated Continuous Lateral Rotation Therapy (CLRT)
The bariatric bed must have a low air loss surface width 40  to 50
The bariatric bed must have a low air loss surface length while inflated 80  to 88
The bariatric bed must have a low air loss surface height while inflated (low air loss model) 8.5
Interested parties are encouraged to respond to this notice via e-mail.  Please submit responses to Joseph Locke, joseph.locke@va.gov.  Please title your e-mail  Response to Sources Sought- Bariatric Bed Frames and Mattresses .  Interested parties are respectfully asked to provide the following information.

Company Name
Company Unique Entity Identification (UEI) number
Company DUNS number
Company Business Size
Company Socioeconomic Category
Is your company a Manufacturer or a Dealer/Distributor
Point of Contact   Name
Point of Contact   E-Mail Address
Point of Contact   Phone Number
Point of Contact   Fax Number
If your company is categorized as a Dealer/Distributor, please provide the (Company Name, UEI, DUNS and Business Size and location) for the Manufacturer, to include the Country of Origin for manufactured end products.
Can you quote product(s) that meets or exceeds all stated salient characteristics?
Do you have any recent / relevant procurement history to document the products being used at VA Hospitals for acute and/or long-term care?  If so, please provide contract numbers.  If history is available for Private Hospitals, please annotate as such.
References to include name, phone number and e-mail address.
Product Evaluation Reports to document customer satisfaction.
Please advise how your company meets the following FAR requirements:
FAR 52.219-14, Limitations on Subcontracting (OCT 2022).
FAR 52.219-33, Nonmanufacturer Rule

## Attachments

- [36C24725Q0421_1.docx](https://sam.gov/api/prod/opps/v3/opportunities/resources/files/ea0a281373e148df8f5805559a66d06e/download)
- [Augusta GA VAMC](https://www.va.gov/augusta-health-care/) (link)

## Open now in the same industry

20 open in NAICS 339113.

- VALVE,BREATHING APP (N0010427QLA17), due November 27, 2026 at 3:30 PM EST. https://sealedbidder.com/opportunities/n0010427qla17
- REAGENT POWDER (N0010426QEJ19), due November 30, 2026 at 3:30 PM EST. https://sealedbidder.com/opportunities/n0010426qej19
- REMOVABLE CONE,SPEC (N0010427QBB84), due November 30, 2026 at 3:30 PM EST. https://sealedbidder.com/opportunities/n0010427qbb84

## Publications

- April 10, 2025: Sources sought, due April 15, 2025 at 10:00 AM EDT. Notice ee039dcfa8f74067953baf73b8c80f29. https://sam.gov/workspace/contract/opp/ee039dcfa8f74067953baf73b8c80f29/view

## Points of contact

- joseph.locke@va.gov, Joseph Locke, Joseph.locke@va.gov, 803-776-4000 x7814

---
Source: SAM.gov Contract Opportunities bulk extract. Confirm deadlines on SAM.gov before responding. Cite https://sealedbidder.com/opportunities/36c24725q0421.