# Richmond Modular Knee Positioner

Canonical: https://sealedbidder.com/opportunities/36c24625q0847

- Solicitation number: 36C24625Q0847
- Notice type: Special notice (first published as sources sought)
- Status: Awarded to Innovative Medical Products, Inc.
- Department: Department of Veterans Affairs
- Contracting office: 246-Network Contracting Office 6 (36C246)
- NAICS: 339112 Surgical and Medical Instrument Manufacturing
- Product or service code: 6515 Medical and Surgical Instruments, Equipment, and Supplies
- Small business size standard: 1,000 employees (13 CFR 121.201, NAICS 339112)
- Who can bid, by location: Anywhere in the US. Location does not limit who can bid: a product shipped to the buyer, research at the performer's facility, or work at the contractor's site. A state shown is the delivery point.
- Location read from: the notice's place of performance: "Product purchase (PSC 6515)"
- County: City of Richmond (FIPS 51760). https://sealedbidder.com/counties/city-of-richmond-va-51760
- City: Richmond. https://sealedbidder.com/cities/richmond-va-5167000
- First posted: June 27, 2025
- Last posted: July 2, 2025
- SAM.gov: https://sam.gov/workspace/contract/opp/587aa2baf16845038803ef352a0fdc42/view

## Reading this notice

- Kind of notice: Special notice. Industry days, draft requirements, program announcements and other notices that are not a solicitation.
- How it ended: Awarded to Innovative Medical Products, Inc. on August 8, 2025, 37 days after offers were due. The contracting office reported 1 offer (not competed).

## Description

The Department of Veterans Affairs, Network Contracting Office (NCO) 6, intends to negotiate a sole source contract with INNOVATIVE MEDICAL PRODUCTS, INC.

This requirement is for Modular Knee Positioners at the Richmond VAMC, 1201 Broad Rock Blvd, Richmond, VA 23249. The machine must meet the following minimum Salient Characteristics:
Knee Positioner   Salient Characteristics
What It Is:
A Precision Knee Positioner is a specialized surgical device used in orthopedic procedures, particularly total knee arthroplasty (TKA) and other knee surgeries. It is designed to stabilize and manipulate the patient's leg during surgery, providing optimal exposure and consistent positioning of the knee joint.
What It Does:
Stabilizes the Leg: Secures the leg and maintains consistent alignment throughout the procedure.
Allows Controlled Movement: Enables surgeons to flex, extend, rotate, or lock the knee in specific positions to access different anatomical structures with precision.
Enhances Surgical Accuracy: Assists in maintaining consistent limb alignment and joint exposure, improving implant placement and surgical outcomes.
Improves Efficiency and Ergonomics: Reduces the need for manual holding by surgical staff, minimizing fatigue and allowing for better workflow in the operating room.
What It Looks Like:
Typically consists of a metal or radiolucent frame mounted to the surgical table.
Features adjustable padded supports or cradles to hold the thigh, calf, and foot securely.
Includes mechanical arms or hinges that allow for controlled movement of the leg in multiple planes.
Some models are equipped with quick-release locks, angle indicators, or motorized components for precision adjustments.
The design is compact, sturdy, and often modular to accommodate different patient anatomies and surgical techniques.
Physical Characteristics:
1. Soft Tissue Protection
The device shall include high-density padding and foam-lined contact surfaces that conform to the patient s anatomy.
The device shall include cushioned leg and femoral support specifically designed to distribute weight evenly to reduce the risk of neurovascular compression.

2. Stability
The device shall contain a multi-point fixation design.
The femoral and distal leg cradles shall be mounted on a rigid base platform that resists lateral movement and rotational drift.
The device shall use a non-slip surface interface and clamping to the surgical table rail.
3. Adjustability
The device allows for multi-axis adjustability, enabling controlled manipulation of flexion, extension, abduction, adduction, and rotation. Range of motion typically spans:
0Â° to 135Â° flexion
Â±20Â° abduction/adduction
Internal/external rotation up to 20Â°
The device shall make adjustments electronically using a footswitch or handheld controller without manual locking or physical readjustment.
4. Locking Mechanism

The positioner shall use a spring-loaded, cam-style locking mechanism that allows for quick engagement and secure hold of the limb at any point along its range of motion.
The locking system shall be actuated via a single-hand or foot-activated release.
The mechanism shall ensure zero drift once engaged and is designed for tool-free operation.

5. Radiolucency
The leg support and cradle system shall be constructed from radiolucent polymer composites, allowing for clear intraoperative fluoroscopic imaging without obstruction.
Imaging shall be able to be taken at 90Â° knee flexion in the axial and lateral planes, and the material composition does not interfere with X-ray or C-arm visibility.
6. Surgical Efficiency
The system shall be designed for no more than 2 3 assembly steps.
The entire apparatus shall be capable of being affixed to the operating table rail via a universal clamp system, with alignment guides and quick-connect components reducing pre-op configuration time.
Thae device shall contain preset angle indicators and sterile draping compatibility.

Period of performance will be for the period 15 October 2024 through 14 October 2025.

Justification for this sole source procurement is in accordance with FAR 13.106-1(b)(2) soliciting from a single source.

This notice of intent is not a request for competitive quotes. No solicitation documents are available, and telephone requests will not be accepted. However, any business that believes it can meet these requirements may give written notice. Supporting evidence must be furnished in sufficient detail to demonstrate the ability to comply with the above requirement. Responses will be considered by the Government, however, a determination by the Government not to compete the proposed acquisition based upon responses of this notice is solely within the discretion of the Government.

NAICS Code 339112/ 1000 employees.

Responses received by 12:00 pm est on 7 July 2025 will be considered by the Government.  Inquiries shall be emailed to stephen.nassan@va.gov. No telephone request will be accepted.

## Attachments

- [36C24625Q0847.docx](https://sam.gov/api/prod/opps/v3/opportunities/resources/files/033debe3c85140e8b659f27d8bd02ed4/download)
- [36C24625Q0847_1.docx](https://sam.gov/api/prod/opps/v3/opportunities/resources/files/9ee6bb3774784aa5b77fff38ee6fba01/download)
- [va.gov](https://www.va.gov/) (link)

## Award on USAspending

- Recipient: Innovative Medical Products, Inc. (UEI KG48A3HJ3V56)
- Contract: 36C24625P1419, purchase order
- Obligated: $31,543.99, current value $31,547
- Competition: Not Competed, 1 offers received
- Link: solicitation number 36C24625Q0847 equals the FPDS solicitation identifier; same awarding office 36C246 (high confidence)
- Record: https://www.usaspending.gov/award/CONT_AWD_36C24625P1419_3600_-NONE-_-NONE-/


## Open now in the same industry

15 open in NAICS 339112.

- Automated External Defibrillator (AED) (W911S227U0200), due November 26, 2026 at 3:15 PM EST. https://sealedbidder.com/opportunities/w911s227u0200
- Vivarium Tunnel Washer (36C24727Q0119), due November 30, 2026 at 8:00 AM EST. https://sealedbidder.com/opportunities/36c24727q0119
- NX EQ Microscope, Light, Operating, Hand/Plastic Surgery (VA-27-00017958) (36C10X27Q0025), due November 30, 2026 at 10:00 AM EST. https://sealedbidder.com/opportunities/36c10x27q0025

## Publications

- June 27, 2025: Sources sought, due July 2, 2025 at 3:00 PM EDT. Notice 2a0e690add31489a9d0cda7761214773. https://sam.gov/workspace/contract/opp/2a0e690add31489a9d0cda7761214773/view
- July 2, 2025: Special notice, due July 7, 2025 at 12:00 PM EDT. Notice 587aa2baf16845038803ef352a0fdc42. https://sam.gov/workspace/contract/opp/587aa2baf16845038803ef352a0fdc42/view

## Points of contact

- Contract Specialist, Stephen Nassan, stephen.nassan@va.gov, 757-251-5118

---
Source: SAM.gov Contract Opportunities bulk extract and USAspending.gov award data. Confirm deadlines on SAM.gov before responding. Cite https://sealedbidder.com/opportunities/36c24625q0847.