4 - QEO Spectrometers
145022-25-0004
National Oceanic and Atmospheric Administration, Department of Commerce NOAA. Analytical Laboratory Instrument Manufacturing.
Response deadline
August 21, 2025 at 5:00 PM EDT
Closed 461 days ago. Posted August 14, 2025. Scheduled to archive September 5, 2025.
Get the next one like this from National Oceanic and Atmospheric Administration by email, the day it is posted.Find my opportunities
This one closed in 2025
62 like it are open now in analytical laboratory instrument manufacturing
- Notice of Intent - Illumina MiSeq i100 series instruments
National Institutes of Health, National Institutes of Health - CC
PresolicitationNAICS 334516Ship to Bethesda, MDNOI-OLAO-CC-27-000836Due in 2 daysNov 27 - Graphite Furnace Atomic Absorption Spectrometer
National Institute of Standards and Technology, Department of Commerce NIST
Sources soughtNAICS 334516Ship to Gaithersburg, MDNIST-MML-FY27-SS01Due in 3 daysNov 28 - Intent to Solicit Only One Source for Six (6) Twinleaf Pulsed Pump Magnetometer (part number: PPM-RS422) and one (1) Twinleaf 8 Channel Synchronization Hub (part number: HUB-RS422-8) manufactured by Twinleaf.
Department of the Army, W6QK ACC-APG Adelphi
SolicitationNAICS 334516Ship to MDW911QX27QA015Due in 6 daysDec 1
All 62 open in analytical laboratory instrument manufacturing
Reading this notice
Worked out from the publications and the award on this page.
- Kind of notice
- Combined synopsis and solicitation. A streamlined commercial-item buy where the notice itself is the solicitation. Glossary
- Who may compete
- Total small business set-aside. Only small businesses may compete. Small, under NAICS 334516 (analytical laboratory instrument manufacturing), means no more than 1,000 employees, counted with affiliates. Open Small business opportunities
- Time to respond
- Bidders had 7 days from the day the solicitation was posted, August 14, 2025, to the deadline.
Size standard from 13 CFR 121.201. The solicitation itself is what binds: check the code and standard it states.
Description
As published on SAM.gov.
ARL conducts applied research into atmospheric chemistry and transport processes, air-surface exchange, and numerical model development and evaluation. To advance this mission, ARL’s ASMD is collaborating with scientists at the Chemical Sciences Lab (CSL) to develop a cutting-edge 4-channel instrument capable of measuring NO, NO2, NOy, O3 in an air aircraft mission in collaboration with scientists at CSL. The instrument is based on the state-of-the-art technique called broadband Cavity Enhanced Absorption Spectroscopy (CEAS). This development is essential for acquiring critical data to better understand the mechanisms of air pollution formation and atmospheric photochemical processes.
The instrument will be developed and then be deployed on the NOAA Twin Otter aircraft in summer 2025 and will subsequently be utilized in AiRMAPS field studies in the years that follow. This advanced system comprises a new set of four spectrometers, which form the core of the 4-channel NO/NO₂/NOy/O₃ instrument. The development of this innovative tool is critical for enabling fast, precise measurements of nitrogen species (NO, NO₂, and NOy) and ozone (O₃). These measurements are essential for advancing our understanding of atmospheric chemistry and air pollution dynamics.
The spectrometers must meet the following requirements to ensure the instrument's performance and accuracy:
- Wavelength range: ~185-1100 nm (user configurations available within this range)
- Optical resolution (w/10 µm slit): 1.20-1.60 nm (FWHM) (preconfigured models)
- Integration time: 8 ms-60 minutes
- Dynamic range: 85000:1 (single scan)
- Signal to Noise Ratio (single scan): 1000:1
- Buffering: stores up to 15,000
- spectra Interfaces: USB; SMA 905; RS-232
- TEC: -40 °C to 50 °C
- Temperature (storage): -30 °C to 70 °C
- Temperature (operation): 0 °C to 55 °C
- Weight: less than 3 lbs
We have identified Ocean Optics, Inc. as the vendor for spectrometers that meet the specified requirements. The selected models are high-sensitivity QEPro spectrometers (Model QEPro-UV250-200), which offer exceptional data collection rates. These spectrometers have been successfully utilized by NOAA/CSL and University of Alabama Huntsville in previous similar developments, demonstrating their reliability and suitability for this application.
Ship to Address
NOAA Air Resources Lab
5830 University Research Court
College Park, MD 20740
Attachments
Files listed on SAM.gov. The ones we hold a copy of download from here; the rest open SAM's download, which needs no account.
- COMBINED SYNOPSIS_SOLICITATION 145022-25-0004.pdf848 KB, August 14, 2025
Publications
Every notice SAM.gov issued under this solicitation number, oldest first. Each is a separate record on SAM.
August 14, 2025
Combined synopsis and solicitation
Due August 21, 2025 at 5:00 PM EDT. SAM.gov, notice 406defa84bda463eaf69ccb0fe7a2666
1 revision on SAM
- August 14, 2025: publish, combined synopsis and solicitation
Points of contact
- Barbara Shifflettbarbara.shifflett@noaa.gov8654054493
- Gabrielle Landgabrielle.Land@noaa.gov865-264-0283