# High-Performance Liquid Chromatography System

Canonical: https://sealedbidder.com/opportunities/1164354

- Solicitation number: 1164354
- Notice type: Sources sought
- Status: Closed. Deadline was March 13, 2026 at 12:00 PM EDT
- Department: Department of Agriculture
- Agency: Agricultural Research Service
- Contracting office: USDA ARS Afm Apd (1232SA)
- NAICS: 334516 Analytical Laboratory Instrument Manufacturing
- Product or service code: 6640 Laboratory Equipment and Supplies
- Small business size standard: 1,000 employees (13 CFR 121.201, NAICS 334516)
- Place of performance: Madison, Wisconsin
- Who can bid, by location: Anywhere in the US. Location does not limit who can bid: a product shipped to the buyer, research at the performer's facility, or work at the contractor's site. A state shown is the delivery point.
- County: Dane County (FIPS 55025). https://sealedbidder.com/counties/dane-county-wi-55025
- City: Madison. https://sealedbidder.com/cities/madison-wi-5548000
- First posted: March 3, 2026
- Last posted: March 3, 2026
- SAM.gov: https://sam.gov/workspace/contract/opp/34d7c0b6f2bc47368f5471be8a4fdf5c/view

## Reading this notice

- Kind of notice: Sources sought. Market research. The agency wants to know who could do the work before it decides how to compete it.
- Who may compete: SAM.gov lists no set-aside on this notice, so it is not reserved for small businesses. Small, under NAICS 334516 (analytical laboratory instrument manufacturing), means no more than 1,000 employees, counted with affiliates.
- What comes next: No solicitation was published under this number after the sources sought of March 3, 2026. An agency sometimes issues the solicitation under a new number.

## Description

Description of Specifications - HPLC system with ECD, including computer, software, and installation:

Minimum requirements for the HPLC system we require:

	The system should have all essential parts of an HPLC, to include: Autosampler, a degasser module, pumps, a column oven and columns, a detector, a controller, and all of the necessary tubing and accessories required to run the system, including the software necessary to connect to the controller and run the system. All major parts of the system (I.E. the pumps, detector, column oven, controller, etc.) should be from the same manufacturer or at least supported (as in troubleshooting support) by the same manufacturer.

	The parts of the system which lie within the flow path which carries the various buffers and samples through the system must not contain metal. No part of the flow path should contain stainless steel and all tubing used should consist of PEEK tubing, a hardy, non-reactive plastic tubing commonly used for various HPLC setups where metal is not possible due to the nature of the assayed samples and/or buffers necessary during analysis. PEEK tubing is required . The system and flow path must also be capable of enduring a wide range of pH, from 1.0-14.0, as many buffers used in our assays are at the extreme ends of the pH scale. Minimum pressure tolerance of at least 20 MPa (2900 psi)

	The system must have multiple pumps (at least 2), or at absolute minimum a single pump capable of generating gradients from multiple inputs.

	 Specific non-detector potions of the system:

		The solvent delivery system should be able to operate in constant flow or constant pressure modes.

		The pumps must have a high flow rate accuracy (+ or - 2%) and precision (0.06%)

		The pumps should come equipped with a suction filter to prevent particulate contamination. As this will be in the flow path, it should be non-metal, ideally ceramic.

		The autosampler must provide temperature control to the sample compartment (between 4-40C minimum)

		The autosampler must have a high injection volume accuracy (+ or &ndash; 1%) and low injection volume variance (<0.3%)

	The HPLC will require a specific type of detector called an electrochemical detector (ECD). The detector should either be manufactured by the same manufacturer as the other major parts, or at least be supported (as in Troubleshooting support) by the same manufacturer as the other major parts. The detector should also come with an Au (gold) electrode, as this is the electrode type needed for sugars analysis.  The system is perferred to be modular and allow the installation of additional detectors in future.

	Any HPLC option considered should include technical installation and initial testing of the system;. The technician performing the installation is required to be a U.S. Citizen.

	Additional System Specifications:

	System Controller

		Supports remote monitoring

		Mobile phases can be purged and baselines checked easily via touch panel.

	Solvent Delivery Pump

		Auto-diagnostics to detect problems during analysis, and an auto-recovery function

		Duel solvent delivery system to reduce space

	Degassing Unit

		Can handle up to 3 degassed solvents

	Autosampler/Plate Changer

		Fast injection speed

		Continuous analysis of hundreds of samples

		High-reproducibility and ultra-low carryover for micro-volume injections

		Automated sample preparation functions

	Column Oven

		Automatic management of column information such as number of injections and pressure during analysis

		Monitoring of column degradation through usage history

	Electrochemical Detector

		Ability to detect carbohydrates using high-performance anion exchange chromatography with pulsed amperometric detection.

There should be a warrantee of at least 1 year, and preferably 2 years or more, for the system.

## Open now in the same industry

55 open in NAICS 334516.

- Incubator Shaker (W911S227U0167), due November 26, 2026 at 11:00 AM EST. https://sealedbidder.com/opportunities/w911s227u0167
- iWorx (W911S227U0166), due November 26, 2026 at 11:00 AM EST. https://sealedbidder.com/opportunities/w911s227u0166
- NANO-AEROSOL HYGROSCOPICITY- CLOUD CONDENSATION ANALYZER (FA860127Q0031), due November 27, 2026 at 11:00 AM EST. https://sealedbidder.com/opportunities/fa860127q0031

## Publications

- March 3, 2026: Sources sought, due March 13, 2026 at 12:00 PM EDT. Notice 34d7c0b6f2bc47368f5471be8a4fdf5c. https://sam.gov/workspace/contract/opp/34d7c0b6f2bc47368f5471be8a4fdf5c/view

## Points of contact

- Maria Thomas, Maria.Thomas@usda.gov

---
Source: SAM.gov Contract Opportunities bulk extract. Confirm deadlines on SAM.gov before responding. Cite https://sealedbidder.com/opportunities/1164354.